WorksheetsMCP HEATER COOLER MAQUET HCU-40 PRE TEST & POST TEST
Total questions: 10
Worksheet time: 4mins
which is the picture of Heater Cooler :
which is the picture of Control Unit HCU-40 :
Which one is'nt a picture Heater Cooler:
Which fluid is suitable for Heater Cooler HCU-40?
Aqua Galon/Refiil
AquaDest
Accu Zur
Infusion Fluid (Sodium Chloride)
What is the tank capacity of Heater Cooler Maquet HCU-40?
28 Liters
30 Liters
25 Liters
34 Liters
What Ampere MCB in the hospital is suitable for Heater Cooler Maquet HCU-40?
16 Ampere (Single Line)
16 Ampere ( 1 Line for Double)
6 Ampere
10 Ampere
If the heater cooler have alarm " water flow too Low/High" what should we do?
Open the machine and chek the flow sensor
Smile and giving money
Check kran, Chek connection and blanket, perform dearing again
chek power cable
If the liquid in the tank is full but in the screen does'nt indicate there is a water(Alarm "Water Tank Empty"), what should we do?
Open the machine, swab the sensor box kit
Take the avometer, measure the resistance, if no resistance change the water
Drain the fluid and change the water sensor
take a selfi photo then send to the doctor
If the perfusion ask you, can we use 2 source temperature with different temperature in the heater cooler HCU-40 running together, then your correct answer is?
Yes, Use Cardioplegic & Patient / Blanket
Yes, Use Patient & Blanket
Smile, giving little joke and dissappear
No, we only have one way
If the Alarm "Tank Overflow! Water level to high", what should we do?
Drain the water from the tank until the cheklist green mark in the screen activied
Drain the water with straw
Drain the water with rucika pipe
Drain the water from the tank until empty, the perform emptying