Worksheetscobas 6000 Day 2 Review
Total questions: 16
Worksheet time: 2hrs 33mins
Which sample containers are specified on cobas 6000?
Ø 13 or 16 x 75 or 100 mm tube
Hitachi cup, microcup, cup on tube
Non-standard tubes Ø 12 to 16 x 73-102 mm
3 false bottom tube (FBT) per system
What could you tell me about a system called e201?
e = Elecsys
2 = rather low throughput
0 = works with core; possibly hybrid
1 = 1st generation
For what are microcups not allowed? Why not?
Microcups are allowed only for sample measurements on the c501 module.
Microcups are allowed only for sample measurements on the e 601 module.
System cannot differentiate with height measurement between standard Hitachi and microcup
Calibration & QC
Functions of the cobas® link.
Provide up-to-date IRD to CU
Provide up-to-date HRD to operator
Storage of backup from CU
Remote-access gateway for support
What does STAT stand for?
Short Turnaround Time
See The A Test
Short Cut A Test
When is the Rack Adapter needed?
For Ø 13 mm tubes a rack adapter is mandatory to prevent the risk of low flyers.
For Ø 13-16 mm tube
Mic Cup
FBT
Which solutions does the ISE probe pipette?
Internal Standard
ISE diluent
Ecotergent (Eco-D)
Sample
What is ICVC and where can we switch it off?
Utility>System>Special functions
Integrated Cassette Volume Check using a pressure sensor
can not switch it off
All c packs which are not filled by Roche, i.e. partner channels and MULTI/CDC c packs
Name the possible root causes for a Duplicate error during c501 calibration.
Foam or bubbles on calibrator material
reagent pipetting problems
old cuvettes/lamp
Dirty water bath
How many measuring points are in a 10 min reaction?
70
57
46
38
What is the maximum throughput of a cobas® 6000 system under ideal conditions
ccc,3000 tests/hrs
cce,2170 tests/hr
cee,1340 tests/hr
eee,510 tests/hr
You are looking for information on cobas 6000.. Name the important documents which you would consult?
OM, Operator Manual
TG,Training Guide
COBI-CD,
Compendium of Background Information
host manual
What is a ghost cassette?
A ghost cassette is a c pack which was unloaded but the system did not register the removal
due to e.g. a gripper problem or red alarm during unloading.
Check the position>place any c pack manually >repeat unloading
a c pack used empty
在什么时候可以手动卸载cobas c501上一盒当前正在使用
的c pack试剂?
在stand by状态下卸载
可以在stand by和operation状态下卸载
可以在任何状态下卸载
在cobas c 501上一个加样cycle time是多久?
18秒
8秒
6秒
关于reagent probe1的压力传感器的作用以下哪种提法是正确的?
如果激活了ICVC功能,所有的c pack试剂盒都启用
reagent probe的LLD功能是由于压力传感器提供的
所有非罗氏的c pack试剂盒都启动ICVC(首次上机)