cobas 6000 Day 2 Review

Total questions: 16

Worksheet time: 2hrs 33mins

Name
Class
Date
1.

Which sample containers are specified on cobas 6000?

a)

Ø 13 or 16 x 75 or 100 mm tube

b)

Hitachi cup, microcup, cup on tube

c)

Non-standard tubes Ø 12 to 16 x 73-102 mm

d)

3 false bottom tube (FBT) per system

2.

What could you tell me about a system called e201?

a)

e = Elecsys

b)

2 = rather low throughput

c)

0 = works with core; possibly hybrid

d)

1 = 1st generation

3.

For what are microcups not allowed? Why not?

a)

Microcups are allowed only for sample measurements on the c501 module.

b)

Microcups are allowed only for sample measurements on the e 601 module.

c)

System cannot differentiate with height measurement between standard Hitachi and microcup

d)

Calibration & QC

4.

Functions of  the cobas® link.

a)

Provide up-to-date IRD to CU

b)

Provide up-to-date HRD to operator

c)

Storage of backup from CU

d)

Remote-access gateway for support

5.

What does STAT stand for? 

a)

Short Turnaround Time

b)

See The A Test

c)

Short Cut A Test

6.

When is the Rack Adapter needed?

a)

For Ø 13 mm tubes a rack adapter is mandatory to prevent the risk of low flyers.

b)

For Ø 13-16 mm tube

c)

Mic Cup

d)

FBT

7.

Which solutions does the ISE probe pipette?

a)

Internal Standard

b)

ISE diluent

c)

Ecotergent (Eco-D)

d)

Sample

8.

What is ICVC and where can we switch it off?

a)

Utility>System>Special functions

b)

Integrated Cassette Volume Check using a pressure sensor

c)

can not switch it off

d)

All c packs which are not filled by Roche, i.e. partner channels and MULTI/CDC c packs

9.

Name the possible root causes for a Duplicate error during c501 calibration.

a)

Foam or bubbles on calibrator material

b)

reagent pipetting problems

c)

old cuvettes/lamp

d)

Dirty water bath

10.

How many measuring points are in a 10 min reaction?

a)

70

b)

57

c)

46

d)

38

11.

What is the maximum throughput of a cobas® 6000 system under ideal conditions

a)

ccc,3000 tests/hrs

b)

cce,2170 tests/hr

c)

cee,1340 tests/hr

d)

eee,510 tests/hr

12.

You are looking for  information on cobas 6000.. Name the important documents which you would consult?

a)

OM, Operator Manual

b)

TG,Training Guide

c)

COBI-CD,

Compendium of Background Information

d)

host manual

13.

What is a ghost cassette?

a)

A ghost cassette is a c pack which was unloaded but the system did not register the removal 

b)

due to e.g. a gripper problem or red alarm during unloading.

c)

Check the position>place any c pack manually >repeat unloading

d)

a c pack used empty

14.

在什么时候可以手动卸载cobas c501上一盒当前正在使用

的c pack试剂?

a)

在stand by状态下卸载

b)

可以在stand by和operation状态下卸载

c)

可以在任何状态下卸载

15.

在cobas c 501上一个加样cycle time是多久?

a)

18秒

b)

8秒

c)

6秒

16.

关于reagent probe1的压力传感器的作用以下哪种提法是正确的?

a)

如果激活了ICVC功能,所有的c pack试剂盒都启用

b)

reagent probe的LLD功能是由于压力传感器提供的

c)

所有非罗氏的c pack试剂盒都启动ICVC(首次上机)