cobas pure OT Day 3 Review

Total questions: 17

Worksheet time: 1hrs 24mins

Name
Class
Date
1.

True or false?

After loading of samples via the STAT port or a

loading lane the barcode reader reads the barcode

of the rack. Then the height sensor detects the

presence and type of standard sample containers.

a)

True

b)

False

2.

What does RFID stand for?

a)

Radiant fluorescence identification

b)

Radio frequency identification

c)

Random frequency identification

d)

Rapid format identification

3.

Which cobas pure integrated solutions reagents

use RFID for identification?

Choose all that apply.

a)

cobas e pack green

b)

CleanCell M / PreClean II M

c)

cobas c pack green

d)

ProCell II M

e)

ISE reagent bottles

4.

Which sizes of cobas e pack green are available?

a)

100 and 300 tests per cobas e pack green

b)

100 and 200 tests per cobas e pack green

c)

200 and 300 tests per cobas e pack green

d)

200 and 400 tests per cobas e pack green

5.

Where on the cobas pure integrated solutions is

sonic wash applied?

a)

On the cobas c 303 analytical unit S1

b)

On the cobas c 303 analytical unit S2

c)

Nowhere

d)

On the cobas e 402 analytical unit reagent probe and on the cobas c 303

analytical unit S1

6.

What triggers the sample probe wash on the cobas

c 303?

Choose all that apply

a)

After 225 pipettings of the sample probe of the cobas c 303 analytical unit

b)

After Samp.S alarm

c)

After Samp.C and Samp.B alarm

d)

After 150 pipettings on the ISE analytical unit

7.

How is the routine rack label 232 displayed in the

user interface?

a)

50232

b)

N50232

c)

70232

d)

60232

8.

What is the correct setting for the daily wash

rack?

a)

Position 1: 1000 μL ISE Cleaning Solution

Position 2: 600 μL ISE Cleaning Solution

Position 3: 600 μL Activator

Position 4: Empty

Position 5: Empty

b)

Position 1: Empty

Position 2: 600 μL ISE Cleaning Solution

Position 3: 600 μL Activator

Position 4: Empty

Position 5: Empty

c)

Position 1: Empty

Position 2: Empty

Position 3: 600 μL Activator

Position 4: Empty

Position 5: Empty

d)

Position 1: Empty

Position 2: Empty

Position 3: Empty

Position 4: Empty

Position 5: Empty

9.

What is the correct setting for the weekly wash

rack?

a)

Position 1: 1000 μL ISE Cleaning Solution

Position 2: 600 μL ISE Cleaning Solution

Position 3: 600 μL Activator

Position 4: Empty

Position 5: Empty

b)

Position 1: Empty

Position 2: 600 μL ISE Cleaning Solution

Position 3: 600 μL Activator

Position 4: Empty

Position 5: Empty

c)

Position 1: Empty

Position 2: Empty

Position 3: 600 μL Activator

Position 4: Empty

Position 5: Empty

d)

Position 1: Empty

Position 2: Empty

Position 3: Empty

Position 4: Empty

Position 5: Empty

10.

What are micro cups used for on the system?

a)

QC

b)

Patient samples

c)

Calibrators

d)

ISE cleaning solution

11.

SOM (self-operating maintenance) is enabled by

default. Who can change the settings for SOM?

a)

Operator

b)

Administrator

c)

Roche Service representative

d)

Patient

12.

True or false?

Administrators can still use the bmserv /nakakojo

service password.

a)

True

b)

False

13.

What is the meaning of "None" in the Rack Ranges

settings?

Choose all that apply

a)

These racks can carry different sample types

b)

The user can leave the sample type open while ordering and define it at a

later time

c)

Sample types of this rack range are sent to cobas pure integrated

solutions by LIS/middleware/infinity

d)

Samples are defined by the barcode of the sample

14.

Under Settings > "Barcode Reading", what does

"sample type encoding in barcode" mean?

a)

The sample type is only accepted if Roche specified barcodes are used

b)

The user can leave the sample type open while ordering and define it later

by adding a barcode number manually

c)

No barcode is required

d)

Sample types are defined by the sample barcode

15.

What are calibration causes?

Choose all that apply.

a)

Update

b)

Calib.New

c)

TimeOut

d)

Changeover

e)

Failed

16.

What needs to be considered if cobas e pack green

are unloaded after use from the cobas e 402

analytical unit?

Choose all that apply.

a)

Unloaded reagent packs need to be specifically registered when the user

intends to load them to a different cobas? pure integrated solutions

analyzer at a later time

b)

The cobas e 402 analytical unit reagents have to be stored according to the

respective Method Sheet

c)

A cobas e pack green which was unloaded cannot be used on the system

after the OBS at the time of reloading is elapsed

d)

Caps of the unloaded cobas e packs green have to be fully closed before

storage

e)

cobas e pack green unloaded from the cobas e 402 analytical unit always

have to be stored in the freezer (-20 C)

17.

Which calibration principles are implemented in the

cobas c 303 analytical unit?

Choose all that apply.

a)

AutoCal

b)

DuoCal

c)

PhotoCal

d)

CopyCal