NFPA 99 Worksheet 6
Authored by Michael Flynn
Professional Development
Used 97+ times

AI Actions
Add similar questions
Adjust reading levels
Convert to real-world scenario
Translate activity
More...
Content View
Student View
25 questions
Show all answers
1.
MULTIPLE CHOICE QUESTION
30 sec • 1 pt
Each station outlet/inlet for medical gas or vacuum shall be ____________.
interchangeable
gas-specific
threaded
quick-coupler
2.
MULTIPLE CHOICE QUESTION
30 sec • 1 pt
The factory installed tubing on station outlets/inlets shall extend no further than ______ from the body of the terminal.
6”
8”
10”
12”
3.
MULTIPLE CHOICE QUESTION
30 sec • 1 pt
Factory installed tubing on station outlets shall be not less than _________ NPS.
1/8 inch
3/16 inch
1/4 inch
5/16 inch
4.
MULTIPLE CHOICE QUESTION
30 sec • 1 pt
Station outlets that operate at a pressure in excess of __________ shall be either D.I.S.S. or be designed to relieve pressure prior to removal of adaptor.
50 psi
60 psi
70 psi
80 psi
5.
MULTIPLE CHOICE QUESTION
30 sec • 1 pt
Station outlets/inlets of manufactured assemblies that employ hose or flexible connections to the piping system that are not fully and immediately accessible, shall use ______________ connectors.
quick coupler
D.I.S.S.
demand check
all of the above
6.
MULTIPLE CHOICE QUESTION
30 sec • 1 pt
The scale range of positive pressure analog indicators shall have the normal reading fall within the ___________ of the total range.
lower third
middle third
top third
none of the above
7.
MULTIPLE CHOICE QUESTION
30 sec • 1 pt
The scale range of vacuum indicators shall be ___________ gauge HgV.
0- 12”
0- 29.9”
0- 30”
0- 30.9”