WorksheetsTraining Quiz AU480 A
Total questions: 31
Worksheet time: 32mins
What is the throughput for Beckman Coulter AU480 Chemistry Analyzer (photometric + ISE optical)?
400
600
800
900
How many type of rack applicable in AU Chemistry analyzer series?
5
4
7
6
How many sample position in STAT Table?
18
22
24
25
How to check Calibration Factor LIst without using the shortcut button or user menu?
Menu > Calibration > Calibration Monitor > Status > Factor List
Men > Calibration > Calibration Detail > Status > Factor List
Menu > Calibration > Calibration History > Status > Factor List
Menu > Calibration Monitor > Calibration Detail > Factor List
How to check a daily Quality Control without using shortcut or user menu?
Menu > QC > QC Monitor > Daily Control
Menu > QC > QC Monitor > Day to day Control
Parameter > QC Monitor > Daily Control
Parameter > QC Monitor > Day to day Control
Which of the following stability period for he ISE reagent once open?
ISE Reference - 2 months / ISE Buffer - 1 Month / ISE Mid Standard - 2 Months
ISE Reference - 2 months / ISE Buffer - 2 Months / ISE Mid Standard - 1 Month
ISE Reference - 3 months / ISE Buffer - 2 Montha / ISE Mid Standard - 1 Month
ISE Reference - 1 month / ISE Buffer - 2 Months / ISE Mid Standard - 1 Month
These condition required for re-calibrate except
Change of detergent
Change Calibrator
The QC is outside SD limit
Change reagent
What are the solution use for wash 1 (W1)?
0.5 % Cleaning Solution
0.5 M HCI
1.0% Cleaning Solution
1.0 M HCl
Which of the priming procedure need to be done after cleaning?
Buffer Prime
Mid/Ref Prime
Total Prime
Prime Bypass
What are the two (2) important things need to be changed every time when performing W2?
Reagent Probe and Sample Probe
Sample syringe and Reagent Syringe
ISE solution and detergent
Pre-Diluent Bottle and W2 Tube in STAT Table
How many cuvettes are there in AU480 Chemistry Analyzer?
86
88
90
80
Which of the following rack function is correct?
Blue - Calibration
Yellow - Reagent Blank
Green - Emergency
Orange - Repeat
White - QC
What is the maximun average photocal reading that user need to change photo lamp?
1.2
0.8
0.9
1.5
How many slot available in the STAT Table?
22 slot for sample, 4 for ISE, 3 for maintenace
22 slot for sample, 4 for ISE, 2 for maintenace
23 slot for sample, 6 for ISE, 3 for maintenace
23 slot for sample, 4 for ISE, 3 for maintenace
What is the correct sequence to perform an Emergency Stop and Restarting system
Turn Off CPU > Push Emergency Stop Button > Turn Off Monitor > Push Power On Button
Push Emergency Stop Button > Push Reset Button > Turn Off CPU > Push Power On Button
Push Emergency Stop Button > Turn Off CPU > Push Reset Button > Push Power On Button
Turn Off CPU > Push Emergency Stop Button > Push Reset Button > Push Power On Button
Which of the following is true about selectivity check procedure?
Perform selectivity check whenever doing priming
Perform selectivity check whenever calibrating ISE
Perform selectivity check daily
Perform selectivity check should be performed after a successful calibration
Which of the following are the factr that affect a calibration result?
Reagent Stability
Calibration Parameter
Quality Control preparation
Parameter set up
Which of the following are the factor that affect a Quality Control result?
Parameter
Calibration
Index
Reagent
What are the solution used for W2?
0.5% Cleaning Solution
1.0% Cleaning Solution
1 M HCl
1 M NaCl
What is the compartment shown in the picture
Cuvette wheel Unit
Reagent Probe
ISE Compartment
Reagent Refrigerator
Name the compartment above
Dry Incubator
STAT Table
Reagent Refrigerator
Cuvette Wheel
Which of the following is the wash nozzle unit?
Pick figure that represent the STAT Table Unit
Which one is the ISE compartment?
Which one is the rack feeder unit
Pick figure that shown the sample probe
Which solution use to perform calibration on Creatinine and Direct Bilirubin
Which of the reagent below use for ISE selectivity check
What are the compartment for above figure
Pre Dilution Bottle
Mix Bar
Wash 1 Bottle
Clean Solution Bottle
The mix bar consist of what type mixer rod
Spiral rod
L-shaped rod
Hook rod
Whisk rod
Home Screen > Stat Status > Sample > Start Entry > Key In sample ID > Choose Test > Entry > Exit > Insert Sample into Stat Table > STAT Status > STAT Start
Above are the sequence use to run sample in STAT Table without LIS. Does the sequence correct
Yes
No