Training Quiz AU480 A

Total questions: 31

Worksheet time: 32mins

Name
Class
Date
1.

What is the throughput for Beckman Coulter AU480 Chemistry Analyzer (photometric + ISE optical)?

a)

400

b)

600

c)

800

d)

900

2.

How many type of rack applicable in AU Chemistry analyzer series?

a)

5

b)

4

c)

7

d)

6

3.

How many sample position in STAT Table?

a)

18

b)

22

c)

24

d)

25

4.

How to check Calibration Factor LIst without using the shortcut button or user menu?

a)

Menu > Calibration > Calibration Monitor > Status > Factor List

b)

Men > Calibration > Calibration Detail > Status > Factor List

c)

Menu > Calibration > Calibration History > Status > Factor List

d)

Menu > Calibration Monitor > Calibration Detail > Factor List

5.

How to check a daily Quality Control without using shortcut or user menu?

a)

Menu > QC > QC Monitor > Daily Control

b)

Menu > QC > QC Monitor > Day to day Control

c)

Parameter > QC Monitor > Daily Control

d)

Parameter > QC Monitor > Day to day Control

6.

Which of the following stability period for he ISE reagent once open?

a)

ISE Reference - 2 months / ISE Buffer - 1 Month / ISE Mid Standard - 2 Months

b)

ISE Reference - 2 months / ISE Buffer - 2 Months / ISE Mid Standard - 1 Month

c)

ISE Reference - 3 months / ISE Buffer - 2 Montha / ISE Mid Standard - 1 Month

d)

ISE Reference - 1 month / ISE Buffer - 2 Months / ISE Mid Standard - 1 Month

7.

These condition required for re-calibrate except

a)

Change of detergent

b)

Change Calibrator

c)

The QC is outside SD limit

d)

Change reagent

8.

What are the solution use for wash 1 (W1)?

a)

0.5 % Cleaning Solution

b)

0.5 M HCI

c)

1.0% Cleaning Solution

d)

1.0 M HCl

9.

Which of the priming procedure need to be done after cleaning?

a)

Buffer Prime

b)

Mid/Ref Prime

c)

Total Prime

d)

Prime Bypass

10.

What are the two (2) important things need to be changed every time when performing W2?

a)

Reagent Probe and Sample Probe

b)

Sample syringe and Reagent Syringe

c)

ISE solution and detergent

d)

Pre-Diluent Bottle and W2 Tube in STAT Table

11.

How many cuvettes are there in AU480 Chemistry Analyzer?

a)

86

b)

88

c)

90

d)

80

12.

Which of the following rack function is correct?

a)

Blue - Calibration

b)

Yellow - Reagent Blank

c)

Green - Emergency

d)

Orange - Repeat

e)

White - QC

13.

What is the maximun average photocal reading that user need to change photo lamp?

a)

1.2

b)

0.8

c)

0.9

d)

1.5

14.

How many slot available in the STAT Table?

a)

22 slot for sample, 4 for ISE, 3 for maintenace

b)

22 slot for sample, 4 for ISE, 2 for maintenace

c)

23 slot for sample, 6 for ISE, 3 for maintenace

d)

23 slot for sample, 4 for ISE, 3 for maintenace

15.

What is the correct sequence to perform an Emergency Stop and Restarting system

a)

Turn Off CPU > Push Emergency Stop Button > Turn Off Monitor > Push Power On Button

b)

Push Emergency Stop Button > Push Reset Button > Turn Off CPU > Push Power On Button

c)

Push Emergency Stop Button > Turn Off CPU > Push Reset Button > Push Power On Button

d)

Turn Off CPU > Push Emergency Stop Button > Push Reset Button > Push Power On Button

16.

Which of the following is true about selectivity check procedure?

a)

Perform selectivity check whenever doing priming

b)

Perform selectivity check whenever calibrating ISE

c)

Perform selectivity check daily

d)

Perform selectivity check should be performed after a successful calibration

17.

Which of the following are the factr that affect a calibration result?

a)

Reagent Stability

b)

Calibration Parameter

c)

Quality Control preparation

d)

Parameter set up

18.

Which of the following are the factor that affect a Quality Control result?

a)

Parameter

b)

Calibration

c)

Index

d)

Reagent

19.

What are the solution used for W2?

a)

0.5% Cleaning Solution

b)

1.0% Cleaning Solution

c)

1 M HCl

d)

1 M NaCl

20.

What is the compartment shown in the picture

a)

Cuvette wheel Unit

b)

Reagent Probe

c)

ISE Compartment

d)

Reagent Refrigerator

21.

Name the compartment above

a)

Dry Incubator

b)

STAT Table

c)

Reagent Refrigerator

d)

Cuvette Wheel

22.

Which of the following is the wash nozzle unit?

a)
b)
c)
d)
23.

Pick figure that represent the STAT Table Unit

a)
b)
c)
d)
24.

Which one is the ISE compartment?

a)
b)
c)
d)
25.

Which one is the rack feeder unit

a)
b)
c)
26.

Pick figure that shown the sample probe

a)
b)
c)
d)
27.

Which solution use to perform calibration on Creatinine and Direct Bilirubin

a)
b)
c)
d)
28.

Which of the reagent below use for ISE selectivity check

a)
b)
c)
d)
29.

What are the compartment for above figure

a)

Pre Dilution Bottle

b)

Mix Bar

c)

Wash 1 Bottle

d)

Clean Solution Bottle

30.

The mix bar consist of what type mixer rod

a)

Spiral rod

b)

L-shaped rod

c)

Hook rod

d)

Whisk rod

31.

Home Screen > Stat Status > Sample > Start Entry > Key In sample ID > Choose Test > Entry > Exit > Insert Sample into Stat Table > STAT Status > STAT Start


Above are the sequence use to run sample in STAT Table without LIS. Does the sequence correct

a)

Yes

b)

No