AU680 TRAINING QUIZ SET B

Total questions: 30

Worksheet time: 15mins

Name
Class
Date
1.

What is the throughput for Beckman Coulter Chemistry Analyzer AU680 (Photometric + ISE optional)?

a)

800

b)

1200

c)

600

d)

1600

2.

How many types of rack applicable in AU Chemistry analyzer series?

a)

4

b)

5

c)

6

d)

7

3.

How many sample position in the STAT Table?

a)

18

b)

22

c)

24

d)

25

4.

How to check the Calibration factor list without using shortcut button or user menu?

a)

Menu > Calibration > Calibration Monitor > Status > Factor List

b)

Menu > Calibration > Calibration Detail > Status > Factor List

c)

Menu > Calibration > Calibration History > Status > Factor List

d)

Menu > Calibration Monitor > Calibration Detail > Factor List

5.

How to check a daily QC without using shortcut button or user menu?

a)

Menu > QC > QC Monitor >Day to Day Chart

b)

Menu > Rack Requisition > QC > QC Monitor > Daily Chart

c)

Menu > QC > QC Monitor > Daily Chart

d)

Menu > Routine > QC > QC specific > Daily Chart

6.

Which of the following stability period for the ISE Reagents once open?

a)

ISE Reference – 2 months / ISE Buffer – 1 month / ISE Mid Standard – 2 months

b)

ISE Reference – 2 months / ISE Buffer – 2 months / ISE Mid Standard – 1 months

c)

ISE Reference – 3 months / ISE Buffer –2 months / ISE Mid Standard – 1 month

d)

ISE Reference – 1 months / ISE Buffer – 2 months / ISE Mid Standard – 1 month

7.

How to run sample in stat table without LIS?

a)

Home Screen > STAT status > Sample>Start Entry >Key in Sample ID>Choose Test/profile>Entry>Exit>Insert Sample into STAT Table>STAT Status>STAT Start

b)

Home Screen > STAT status > select sample > test requisition >Demographic > entry>exit> STAT check > sample assignment > STAT start

c)

Home Screen > test requisition > select sample > STAT status >Demographic > entry>exit> STAT check > sample assignment > STAT start

d)

Home Screen > Demographic > select sample > test requisition >STAT status > entry>exit> STAT check > sample assignment > STAT start

8.

These condition require for re-calibrate except?

a)

Change detergent

b)

Change Calibrator

c)

(The QC is outside SD limit

d)

Change reagent

9.

What is solution used for W1?

a)

2.0% Cleaning Solution

b)

0.5 M HCl

c)

2.0 % detergent

d)

1.0 M HCl

10.

How many samples can continuously run on rack holder for AU680?

a)

150 Sample

b)

200 sample

c)

100 sample

d)

80 sample

11.

What are the two (2) important items need to be changed every time when performing W2?

a)

Reagent Probe and Sample Probe

b)

Sample Syringe and Reagent Syringe

c)

ISE solution and Detergent

d)

Pre Diluents Bottles and Det-1 ( 3 bottles)

12.

How many cuvettes are there in AU680 Chemistry Analyzer?

a)

80

b)

153

c)

165

d)

145

13.

Which are following rack function is correct?

a)
b)
c)
d)
14.

What is the maximum average photocal reading that user needs to change photo lamp?

a)

0.1

b)

3.0

c)

2.0

d)

1.5

15.

How many slot available in the STAT table?

a)

22 slot for sample, 4 for ISE, 1 for Maintenance, 2 for Blank

b)

22 slot for sample, 4 for ISE, 2 for Maintenance

c)

23 slot for sample, 6 for ISE, 3 for Maintenance

d)

23 slot for sample, 6 for ISE, 3 for Maintenance

16.

What is the correct sequence to perform an Emergency Stop and Restarting system?

a)

Turn Off CPU – Push Emergency Stop Button – Turn Off Monitor – Push Power ON Button

b)

Push Emergency Stop Button – Push Reset Button – Turn Off CPU – Push Power ON Button

c)

Push Emergency Stop Button – Turn off CPU – Push Reset Button – Push Power ON Button

d)

Turn Off CPU – Push Emergency Stop Button – Push Reset Button – Push Power ON Button

17.

Which are the following is true about selectivity check procedure?

a)

Perform selectivity check whenever doing priming

b)

Perform selectivity check whenever calibrating ISE

c)

Perform selectivity check daily

d)

Perform selectivity check should be performed after a successful calibration

18.

Which of the following are the factors that affect a calibration result? Hint: More than 1 answer option

a)

Reagent Stability

b)

Calibrator preparation

c)

Quality control preparation

d)

Parameter Set up

19.

Which of the following are the factors that affect a Quality Control result?


I. Parameter

II. Calibration

III. Index

IV. Reagent

a)

I and II

b)

I and III

c)

I, II and III

d)

I, II and IV

20.

What are solutions used for W2?


I. 0.5% Cleaning Solution

II. 1.0% Cleaning Solution

III. 1M HCL

IV. 1M NaCL

a)

I and III

b)

II and III

c)

I and IV

d)

II and IV

21.

Please check the following match which one is correct?


A- Reagent Compartment

B- ISE Compartment

H- STAT Compartment

I- Mixer

J- Sample Probe

a)

A, B and J

b)

B, H and I

c)

J, I and H

d)

H, A and J

22.

Please choose the correct sequence for sample processing?

a)

1>3>4>2>7>9>8>6>5

b)

1>2>3>4>5>6>7>8>9

c)

9>8>7>6>5>4>3>2>1

d)

2>4>6>8>1>3>7>9>5

23.

Please choose the correct match. Hint: More than 1 answer option

a)

A- Home

b)

C- Menu List

c)

E- Shutdown

d)

D- Stop/ Standby

e)

B- User Menu

24.

Please choose the correct match. Hint: More than 1 answer option

a)

61- Reagent Probe

b)

62- Reagent Probe

c)

63- Reagent Probe

d)

64- Sample Probe

e)

65- Reagent Probe

25.

Where should you go to view the system temperatures? Circle the correct answer

a)

Start Condition

b)

Analyzer Status

c)

Analyzer Maintenance

d)

Sample Status

26.

(ISE Labs only) Identify the reagents used by the ISE module Hint: More than 1 answer option

a)

ISE Buffer Solution

b)

ISE Mid Standard Solution

c)

Cleaning Solution

d)

ISE Reference Solution

27.

(ISE Labs only) TRUE or FALSE?

You need to press the green TABLE ROTATION/DIAG button on the analyzer to start the ISE Cleaning procedure.

a)

TRUE

b)

FALSE

28.

What should you select to identify the Onboard Stability of reagents?

a)

Reagent Inventory

b)

Reagent Consumption

c)

Reagent Check

d)

Detail Tab

29.

Identify the options available when performing a reagent check. Hint: More than 1 answer option

a)

Check all positions

b)

Check specified positions

c)

Check changed positions

d)

Reset Only

30.

Choose the statement that is false when loading racks in the rack feeder

a)

Caps should be removed from all sample containers

b)

The rack should be placed in the rack feeder with the rack bar code facing the back of the analyzer

c)

Manually programmed samples require a bar code if the analyzer is set for Barcode analysis

d)

When calibrating, always place the yellow racks in front of the blue rack