Module 3: Microtomy

Total questions: 21

Worksheet time: 11mins

Name
Class
Date
1.

microtome most commonly used in histo lab

a)

clinical freezing

b)

sliding microtome

c)

ultra microtome

d)

rotary micromtome

2.

microtome which the block is stationary and a knife is moved horizontally

a)

clinical freezing

b)

rotary

c)

sliding

3.

which microtome moves the block up and down ?

a)

sliding

b)

clinical freezing

c)

cryostat

d)

rotary

4.

which microtome sections large paraffin blocks?

a)

rotary

b)

clinical freezing

c)

vibrating

d)

sliding

5.

which microtome cuts fresh, frozen tissue

a)

rotary

b)

sliding

c)

clinical freezing

6.

which microtome clamps to the tabletop?

a)

ultramicrotome

b)

sliding

c)

clinical freezing

7.

angle formed by the intersection of the block face and adjacent cutting facet of knife.

a)

bevel angle

b)

clearance angle

8.

angle that is usually 28-32 degrees

a)

bevel angle

b)

clearance angle

c)

wedge angle

9.

angle that is set to 3-8 degrees

a)

bevel angle

b)

clearance angle

c)

wedge angle

10.

angle made when 2 cutting facets intersect

a)

bevel angle

b)

clearance angle

c)

wedge angle

11.

angle between cutting facet and block of tissue.

a)

wedge angle

b)

clearance angle

c)

bevel angle

12.

angle that is set at 15 degrees

a)

clearance angle

b)

wedge angle

c)

bevel angle

13.

this blade is used for dense tissue

a)

low profile

b)

high profile

14.

this blade is used for routine sectioning

a)

high profile

b)

low profile

15.

what is most likely the cause for crooked ribbons?

a)

insufficient dehydration or clearing

b)

too little blade tilt

c)

top and bottom edges not parallel

d)

dull blade

16.

what is most likely the cause for washboarding?

a)

dull blade

b)

too much blade tilt

c)

too little blade tilt

d)

block not evenly chilled

17.

what is most likely the cause for vertical scratches in the sections?

a)

defect in blade edge

b)

dull blade

c)

insufficient dehydration during processing

d)

paraffin too soft

18.

most likely the cause for failure of a ribbon to form.

a)

too little blade tilt

b)

too much blade tilt

c)

hard tissue

d)

overfixed tissue

19.

how to resolve issue of section lifting from blade as block is raised?

a)

change room temp

b)

decrease blade tilt

c)

increase blade tilt

d)

replace blade

20.

solution for block face unevenly sectioned

a)

face less aggressively

b)

increase blade tilt

c)

choose paraffin with lower melting point

d)

realign microtome

21.

solution for holes in sections

a)

realign microtome

b)

ensure block is evenly chilled

c)

increase blade tilt

d)

face less aggressively