Hemodialysis Project 3
Authored by NABIHAH A19EB0027
Other
University
Used 5+ times

AI Actions
Add similar questions
Adjust reading levels
Convert to real-world scenario
Translate activity
More...
Content View
Student View
6 questions
Show all answers
1.
MULTIPLE SELECT QUESTION
45 sec • 1 pt
Audible and visual alarms mus be activate when the blood circuit air protection detect?
Series Microbubbles > 1.5ml/10sec
Individual Bolus Air > 1.5ml/30sec
Individual Bolus Air > 1ml
Series Microbubbles > 1.5ml/30sec
2.
MULTIPLE CHOICE QUESTION
30 sec • 1 pt
What is the volume when the alarms activated if the blood leak occur?
0.25mL/sec
0.35mL/min
0.25mL/min
0.35ml/sec
3.
MULTIPLE CHOICE QUESTION
30 sec • 1 pt
What are the two test requirements for hemodialyzer clearance?
Kt/V and URT
URR and K
URR and Kt/V
URT and K
4.
MULTIPLE CHOICE QUESTION
30 sec • 1 pt
The conductivity of dialysate in a functional hemodialysis machine should have an overall functionality of
±10%
±15%
±5%
±7%
5.
MULTIPLE CHOICE QUESTION
30 sec • 1 pt
The temperature of the dialysate when __ °C, protein denaturation and hemolysis occur.
33
43
46
36
6.
MULTIPLE CHOICE QUESTION
30 sec • 1 pt
For manufacturer-specified blood and filtrate flow rates, the blood inlet and outlet concentrations are measured during steady state.
True
False