1.

System inspections shall be performed prior to concealing piping distribution systems in ________.

a)

walls

b)

ceilings

c)

chases

d)

any of the listed locations

2.

Which of the following does NFPA 99 consider NOT an acceptable qualification for an individual performing maintenance on a Category 1 medical gas system?

a)

ASSE 6010

b)

ASSE 6030

c)

ASSE 6040

d)

Documented training program

3.

Master alarm panels shall NOT communicate ___________ to the alarm initiating devices they monitor.

a)

directly

b)

indirectly

c)

directly or indirectly

d)

there is no specification in NFPA 99

4.

Threaded joints shall be made up with ________________ recommended for oxygen service.

a)

polyvinylchloride tape

b)

potlytetrafluoroehtylene tape

c)

polypropylene tape

d)

polyimide tape

5.

Receivers for medical air shall be equipped with ____________.

a)

Relief valves

b)

Automatic and manual drains

c)

Sight glass

d)

All of the equipment listed is required

6.

The cryogenic fluid central supply system shall have a local signal that visibly indicates the operating status, and an indicator at all master alarms, when or at a predetermined set point before the main supply falls to an average day’s supply indicating ____________.

a)

system failure

b)

maintenance needed

c)

low contents

d)

secondary in use

7.

The master alarm system shall include an indication when the pressure in the main line increases or decreases _______ % from normal oprating pressure.

a)

10

b)

20

c)

30

d)

50

8.

Which of the following is not prohibited, throughout medical gas and vacuum distribution pipeline systems?

a)

Flared and compression type connections

b)

Memory metal fittings

c)

Removable and nonremovable push-fit fittings

d)

Pipe-crimping tools used to permanently stop flow of piping

9.

Brazed joints shall be made using a brazing alloy that exhibits a melting temperature in excess of ________, to retain the integrity of the piping system in the event of fire.

a)

800º F

b)

1000º F

c)

1200º F

d)

1500º F

10.

Natural ventilation shall have the opening(s) located ________ and located to ensure cross ventilation.

a)

within 1 ft. of the floor

b)

within 1 ft. of the ceiling

c)

within 1 ft. of the floor and within 1 ft. of the ceiling

d)

there are no guidelines regarding placement of the opening(s)

11.

Which of these is NOT an NFPA 99 requirement for the labels on patient medical gas piping

at standard pressure?

a)

Labeled using stenciling or adhesive markers

b)

The name or chemical symbol of the gas

c)

The operating pressure

d)

The system color

12.

An oxygen outlet having a minimum flow of 3.5 SCFM, shall have a pressure drop of not more than _________ for the operational flow pressure drop test.

a)

5 psi

b)

10 psi

c)

15 psi

d)

20 psi

13.

Which of the following systems shall not be permitted to be located in an outdoor enclosure?

a)

Manifolds for cryogenic liquid containers

b)

Manifolds for gas cylinders

c)

Medical air compressor sources

d)

Bulk cryogenic liquid systems

14.

The final connection between new piping and an existing system, may be tested with the system gas and must be checked with a leak detectant that does Not contain which of these agents?

a)

Chlorine

b)

Acetone

c)

Freon

d)

Ammonia

15.

Those sections or portions of the piping system, that serve a room or group of rooms on the same story of the facility. This is the definition of which of the following?

a)

Main lines

b)

Branch (lateral) lines

c)

Risers

d)

Vent lines

16.

Carbon dioxide cylinders shall NOT reach temperatures greater than how many degrees Fahrenheit?

a)

110

b)

115

c)

120

d)

125

17.

Medical air compressor intake shall be located outdoors a minimum distance of __ above the ground level.

a)

10 ft

b)

15 ft

c)

20 ft

d)

25 ft

18.

An existing system that is not in strict compliance with provisions of this code shall be permitted to be continued in use, as long as the __ has determined that such use does not constitute a distinct hazard to life.

a)

Engineer

b)

Architect

c)

Authority having jurisdiction

d)

Nursing staff

19.

If underground piping is protected by a conduit, cover or other enclosure it shall be ___________ and not retain groundwater in prolonged contact with the pipe.

a)

self-leveling

b)

self-labeling

c)

self-filling

d)

none that are listed are correct

20.

Which of the following is required of the check valves in an EOSC?

a)

Both valves are located downstream from the tee connection

b)

One valve is located downstream of the main line shutoff valve

c)

One valve is located upstream of the EOSC

d)

Both valves are located downstream from the tee connection, and one valve is located downstream of the main line shutoff valve

21.

Factory installed inlet tubes on station outlets shall not be less than ____ with 0.3 minimum inside diameter.

a)

NPS 1/8, 1/4 O.D.

b)

NPS 1/4, 3/8 O.D.

c)

NPS 3/8, 1/2 O.D.

d)

NPS 1/2, 5/8 O.D.

22.

The cryogenic fluid central supply system shall have a local signal that visibly indicates the operating status, and an indicator at all master alarms, when or at a predetermined set point before the reserve supply begins to supply the system indicating ____________.

a)

reserve in use

b)

maintenance needed

c)

low contents

d)

secondary in use

23.

Where initiating devices are remote from the building and the wiring is to run in compliance with NFPA 70, a ___________ set of wires shall be permitted to be run underground for each signal.

a)

Single

b)

Double

c)

Triple

d)

Quadruple

24.

Medical air source systems shall be controlled to ensure continuous supply, and shall have a(n) __________ restart function, such that the supply of medical air will resume normally after power interruption without manual intervention.

a)

automatic

b)

manual

c)

local

d)

remote

25.

For an oxygen central supply system using concentrators, the system shall be tested for __________.

a)

purity of the air

b)

purity of the oxygen

c)

purity of the vacuum source

d)

all of the tests listed shall be performed

26.

Which type of medical gas has an oxygen concentration > or = 90% and < or = 96%?

a)

Medical air USP

b)

Oxygen USP

c)

Oxygen enriched USP

d)

Oxygen 93 USP

27.

Non-mandated fundamentals and recommended guidelines for medical gas and vacuum systems, shall be located in the __________ of the NFPA 99 code or the ASSE 6000 standards.

a)

first or second chapters

b)

annexes or appendixes

c)

qualifications for performance knowledge

d)

scope or scheme

28.

Installers of medical gas & vacuum systems shall NOT use their certifications to oversee installation _____.

a)

by non-certified personnel

b)

of health care personnel who meet ASSE 6010 requirements

c)

when the installing contractor is certified

d)

when the job foreman is certified

29.

Which of the following is NOT a requirement for a compressed gas reserve supply system when serving as part of a cryogenic fluid central supply system?

a)

A check valve to prevent backflow into the system

b)

At least 3 cylinder connection points

c)

Line pressure alarm switch/sensor located between cylinders and regulator

d)

Relief valve/vent located between cylinders and regulator

30.

Only one service valve is required for each branch off the riser, unless the number of zone valves on that branch exceeds ______ valves.

a)

4

b)

6

c)

8

d)

there is no maximum

31.

Who shall be responsible for developing an emergency plan in the event of medical air loss,

at a Category 2 facility?

a)

Head of maintenance

b)

Facility Staff

c)

Authority having jurisdiction

d)

Chief engineer

32.

Requirements for installing corrugated medical tubing, include all of the following EXCEPT:

a)

Fabricated from copper alloy

b)

Identified by the manufacturer as suitable for oxygen a minimum of every 3 feet

c)

Have a minimum wall thickness of schedule 40

d)

Supports shall be in accordance with manufacturer’s instructions

33.

Refer to diagram 3. What is the minimum distance requirement at letter B?

a)

10 feet

b)

15 feet

c)

20 feet

d)

25 feet

34.

In a Category 1 facility, which of the following is correct regarding the standing vacuum test

for vacuum piping?

a)

During the test the source of the test vacuum shall be disconnected from the piping system

b)

Leakage attributed to specific changes in ambient temperature over 0.5 percent of the standing pressure may be allowed

c)

Standing vacuum test is performed after successful completion of an initial pressure test of positive pressure and vacuum systems

d)

All of the answers apply

35.

Heating, cooling ventilating a process systems serving spaces or providing healthcare functions, shall be provided in accordance with __________.

a)

NFPA 55

b)

ASSE 6000

c)

NFPA 99

d)

ASHRAE 170

36.

For the purposes of this code, an oxygen enriched atmosphere is an atmosphere where the total concentration of oxygen exceeds ________ by volume.

a)

23.5%

b)

23.6%

c)

18.5%

d)

93.5%

37.

The medical air dew point high alarm from each compressor site will indicate at the master alarm panels, when the line pressure dew point is greater than ________.

a)

32º F

b)

35º F

c)

22º F

d)

40º F

38.

The scale range of vacuum indicators shall be ______ in. gauge HgV

a)

0-15

b)

12-15

c)

0-30

d)

15-30

39.

Refer to diagram 4. The item #14 would best be described as a(n) __________ valve.

a)

In-line shutoff

b)

Riser

c)

Service

d)

Main line

40.

Certified Installers of medical gas and vacuum systems shall use their certification to install __________.

a)

systems the meet the requirements of NFPA 99

b)

systems the meet the requirements of NFPA 70

c)

systems as the installing verifier

d)

systems as the installing inspector

41.

A gauge reading exceeding 200 psi in a piping system is defined as which type of pressure?

a)

Absolute

b)

Gauge

c)

High

d)

Positive

42.

Power for master and area alarms shall be from the ___________ as described in chapter 6.

a)

secondary electrical system

b)

essential electrical system

c)

primary electrical system

d)

emergency electrical system

43.

If WAGD is produced by the medical-surgical vacuum source, the total concentration of oxygen shall be maintained below ______ percent.

a)

23.6

b)

25.6

c)

21.6

d)

19.6

44.

An oxygen concentrator will have an alarm if the oxygen concentration lower than ___ percent is observed.

a)

90

b)

91

c)

93

d)

95

45.

Area alarms shall monitor __________ in the area served.

a)

High pressure in positive pressure medical gas lines

b)

Low pressure in positive pressure medical gas lines

c)

Low pressure in medical-surgical vacuum lines

d)

High and Low pressure in positive pressure medical gas lines and Low pressure in medical-surgical vacuum lines

46.

Refer to diagram 4. The item #12 would best be described as a(n) __________.

a)

Master alarm sensor

b)

Area alarm sensor

c)

Local alarm sensor

d)

Zone valve alarm sensor

47.

Identify the following definition: A drug-induced depression of consciousness during which patients cannot be easily aroused, but respond purposefully to repeated or painful stimulation.

a)

General Anesthesia

b)

Deep Sedation

c)

Moderate Sedation

d)

Minimal Sedation

48.

What is a permitted use of the medical-surgical vacuum system?

a)

Liquid and debris disposal

b)

Non-medical or Non-surgical applications

c)

Vacuum steam condensate return

d)

Waste anesthetic gas disposal

49.

Copper tube hangers for medical gas and vacuum distribution piping shall be __________.

a)

Supported from the building structure

b)

Sized for copper tube

c)

Plastic coated in potentially damp locations

d)

All that are listed are required

50.

Pipe labels shall be located at a maximum interval of how many feet?

a)

10

b)

15

c)

20

d)

25

51.

A wireless communication device is employed in a medical gas or vacuum system. Which type of supervision is required for this device?

a)

A visual indicator shall be located in the maintenance/ engineering department

b)

The device shall be located at a nurse’s station

c)

It will be supervised such that failure of communication shall initiate an alarm

d)

The device shall be installed at the medical gas or vacuum source

52.

Which of the following is NOT a requirement for station inlets/outlets?

a)

Each outlet/inlet shall be legibly identified

b)

Station outlets/inlets shall be gas specific, whether outlet/inlet is threaded or is a noninterchangeable quick coupler

c)

Station outlets shall not be permitted to use nonstandard operating pressures

d)

The use of common parts such as springs, O-rings, fasteners, seals, and shutoff poppets shall be permitted

53.

Hypobaric facilities area a building or structure used to house hypobaric chambers for medical applications and procedures, at pressures __________ normal atmospheric pressure.

a)

above

b)

below

c)

above and below

d)

equal to

54.

In central supply systems using gas cylinders, each header shall have a ________ valve to prevent backflow and to permit service to the header.

a)

Check

b)

Bypass

c)

Ball

d)

Relief

55.

Liquid ring compressor systems shall have a reserve medical air standby header (1HR) or a _____ a backup.

a)

Condensate trap

b)

Manifold

c)

Compressor

d)

Bulk system

56.

Dryers, filters, and regulators shall be at least __________, and each component shall be sized to serve the peak calculated demand, with the largest of each component out of service.

a)

simplexed

b)

duplexed

c)

triplexed

d)

quadruplexed

57.

The master alarm system shall include an indication from each compressor site, when the instrument air dew point is greater than _______° F.

a)

-22

b)

-32

c)

-35

d)

-40

58.

The instrument air dew point high alarm from each compressor site will indicate at the master alarm panels, when the line pressure dew point is greater than ________.

a)

-39º F

b)

-22º F

c)

-35º F

d)

-40º F

59.

Which of the following does NFPA 99 consider an acceptable qualification for an individual performing maintenance on a Category 1 medical gas system?

a)

6020 or 6030

b)

6020 or 6010

c)

6010 or 6040

d)

6040 or 6030

60.

Who shall witness the installer-performed standing vacuum test?

a)

Installation contractor superintendent

b)

Medical gas instructor

c)

The authority having jurisdiction, or its designee

d)

A witness is not required

61.

What is the minimum required footage between medical-surgical vacuum exhaust & the air intake or other openings in the building?

a)

12'

b)

15'

c)

20'

d)

25'

62.

Joints throughout the medical gas system, shall be brazed within _______ hours of assembly.

a)

4

b)

8

c)

12

d)

24

63.

Hyperbaric facilities area a building or structure used to house hyperbaric chambers for medical applications and procedures, at pressures __________ normal atmospheric pressure.

a)

above

b)

below

c)

above and below

d)

equal to

64.

Fittings for medical gas systems shall be ____________ capillary fittings complying with ASME B16.22.

a)

wrought copper

b)

cast copper

c)

annealed copper

d)

dwv copper

65.

Which of these is an NFPA 99 requirement for labeling of pipes in a medical gas system?

a)

Color Coding

b)

Fixed label width

c)

Label shall wrap completely around the pipe

d)

Reflective label

66.

The facility shall be permitted to use any inspection procedure(s) or testing methods, established through its own _______________.

a)

trial and error

b)

risk assessment

c)

authority having jurisdiction

d)

authorized facility representative

67.

Pressure/vacuum indicators shall be required at which of the following locations?

a)

Adjacent to a master alarm initiating device

b)

At or in the area alarm panel

c)

On the outlet/inlet side of zone valve

d)

All of the locations listed are required

68.

Design, installation, and operation of medical support gas systems shall consider all hazards involved with pressurized gas, and additional hazards associated with ________________.

a)

drier gases

b)

moister gases

c)

flammable gases

d)

elevated pressure gases

69.

In-line check valves shall have which of the following?

a)

Minimum NPS 3/4 threaded connections

b)

Threaded purge points of 1/8 in. NPT

c)

Required to be removed to be serviced

d)

Threaded purge points of 1/4 in. NPT

70.

Instrument air systems shall be permitted to interconnect with the medical air system

provided the system has ___________.

a)

interconnection is prohibited

b)

at least two or more compressors

c)

been approved by the manufacturer

d)

separate source valves for each system

71.

Medical Gas Rail (MGR) assemblies shall be made of materials with a melting point of at least ______° F.

a)

500

b)

1000

c)

1500

d)

2000

72.

Medical air filters shall be sized for 100% of the system peak calculated demand, and be rated for a minimum of ______ efficiency at ______ micron or greater.

a)

95% @ .5

b)

98% @ .1

c)

100% @ .1

d)

None of the listed requirements are correct

73.

WAGD networks shall provide a WAGD inlet in all locations where _______ is intended to be administered.

a)

Oxygen

b)

Nitrous Oxide and other anesthetics

c)

Nitrogen

d)

Medical Air

74.

What sort of treatment shall be performed to cut ends of a tube BEFORE brazing, for piping medical gas?

a)

An argon purge must be done

b)

Cut ends must be rolled smooth or deburred

c)

Extensions need to be added for brazing

d)

Treatment is NOT required, they may be brazed “as is”

75.

Where initiating devices are remote from the building and the wiring is to run in compliance with NFPA 70, _________ single wires shall be permitted to be run underground for each signal.

a)

1

b)

2

c)

3

d)

4

76.

Instrument air sources shall have either ________________ or ____________________.

a)

two compressors, two standby headers

b)

two compressors, one compressor and a standby header

c)

two standby headers, one standby header and an emergency header

d)

two compressors, one compressor and an emergency header

77.

All valves except zone valve box assemblies shall be ____________.

a)

Located in a secure area

b)

Locked or latched in the operating position

c)

Labeled indicating contents and location served

d)

All of the choices are required

78.

Natural ventilation shall consist of two non-closable louvered openings, each having an aggregate free opening of 24 square in. for each 1000 cubic ft. of the gas designed to be stored in the space. In no case shall the louvered opening be less than _____ square in.

a)

24

b)

48

c)

72

d)

96

79.

In a non-anesthetizing area, on which side of the zone valve shall the area alarm

pressure/vacuum sensor be located?

a)

Station outlet/inlet side

b)

Source side

c)

Upstream side

d)

NFPA 99 does not specify a requirement

80.

Medical Air reconstituted from Oxygen USP and nitrogen NF, produced using a proportioning system:

a)

Shall produce medical air with an oxygen content of 99.5%

b)

Shall produce medical air with an oxygen content of 91%

c)

Shall produce medical air with an oxygen content of 19.5% to 23.5%

d)

Shall produce medical air with an oxygen content of 23.5% to 50%

81.

Which of the following is labeled incorrectly?

a)

Oxygen with green background and white text

b)

Oxygen with white background and green text

c)

Medical air with yellow background and black text

d)

Medical air with white background and black text

82.

Oxygen concentrator supply units output shall have < or equal to 1mg/cubic meter of permanent particulates, sized ____ micron or larger at normal atmospheric pressure.

a)

.01

b)

.1

c)

1

d)

10

83.

The Medical-Surgical Vacuum exhaust shall be located where prevailing winds, adjacent buildings, topography or other influences, will divert the exhaust into ________ areas.

a)

outdoor

b)

occupied

c)

unoccupied

d)

confined

84.

Final line pressure controls shall be sized to ________ of the peak calculated demand.

a)

50%

b)

100%

c)

150%

d)

Not specified

85.

Refer to Diagram #4, identify #13.

a)

In-line valve

b)

Main control valve

c)

Service valve

d)

Zone valve box

86.

Manifolds for gas cylinders shall actuate a(n) __________ signal and shall activate an indicator at all __________ alarm panels, to indicate when or just before changeover from one header to the other.

a)

Local - Master

b)

Master - Area

c)

Master - Local

d)

Area - Master

87.

In an Emergency Oxygen Supply Connection, the check valves are required to be constructed of which of the following materials?

a)

Brass or bronze

b)

Brass, bronze or stainless steel

c)

NFPA 99 does Not establish any material requirements for these check valves

d)

Materials per the EOSC manufacturer’s recommendations

88.

The removal of components within a source system for repair and re-installation or for replacement, shall be treated as new work for the purposes of testing, whenever such work involves which of the following?

a)

Air quality monitoring

b)

Calibration of alarm sensors

c)

Cutting and/or brazing new piping

d)

System control sensors

89.

With the approval of the authority having jurisdiction, in what areas shall the shutoff valves be permitted to be secured to prevent inappropriate access?

a)

Maternity

b)

Psychiatric or pediatric

c)

Anesthetizing

d)

Critical care

90.

Inlet filtration of the medical-surgical vacuum central supply systems shall contain all EXCEPT which of the following essentials?

a)

Filters shall be efficient to 99.97 % HEPA

b)

Filtration shall be simplex

c)

Filtration shall be on the patient side of the vacuum pump

d)

Filtration shall be sized to a minimum of 100% of peak calculated demand

91.

When designing and constructing central supply systems, the storage shall __________.

a)

Have lockable doors or gates

b)

Secure all cylinders

c)

Have a minimum 1-hour fire resistance rating

d)

All of the listed requirements apply

92.

Joints shall be brazed within _______ hours after surfaces are cleaned for brazing.

a)

1

b)

4

c)

8

d)

27

93.

Hard-drawn seamless copper in medical gas systems shall be type L, except when pressures exceed 185 psi, ____________ shall be used.

a)

type K

b)

type M

c)

type DWV

d)

type ACR

94.

A schedule for maintaining medical-surgical inlets within a facility, would include all of the following EXCEPT:

a)

Risk assessment of the facility

b)

Maintenance personnel certified to ASSE 6020 requirements

c)

Recommendations of the original equipment manufacturer

d)

Recommendations as required by the authority having jurisdiction

95.

A central supply system with a storage capacity of more than 3200 pounds at normal temperature and pressure, best describes which system?

a)

Bulk Cryogenic Fluid Central Supply System

b)

Bulk Nitrous Oxide Central Supply System

c)

Cylinder System

d)

Micro Bulk Cryogenic Fluid Central Supply system

96.

To facilitate service, testing or replacement, all pressure switches and mainline gauges downstream of the source valve shall be provided with a __________.

a)

Check valve

b)

Shutoff valve

c)

Demand check fitting

d)

D.I.S.S fitting

97.

Verification tests shall only be performed after all of the _____________ performed tests.

a)

installer

b)

inspector

c)

responsible facility authority

d)

AHJ

98.

Pipe labels shall be located at a maximum interval of how many feet?

a)

10

b)

15

c)

20

d)

25

99.

Immediately outside each Category 1 space and anesthetizing location, there shall be located a(n) ________ for each medical gas, vacuum and WAGD system.

a)

Zone valve

b)

Fire extinguisher

c)

Relief valve

d)

Area alarm

100.

Which of the following fittings is NOT permitted for joining vacuum distribution pipeline systems?

a)

Axially swaged

b)

Non-removable push-fit

c)

Welded

d)

Memory metal