DxC 700 AU

Total questions: 30

Worksheet time: 3600secs

Name
Class
Date
1.

1. What is throughput for Beckman Coulter Chemistry Analyzer DxC 700 AU (photometric + ISE optional)?

a)

800

b)

1200

c)

1000

d)

900

2.

How many R1 Reagent can be loaded into R1 Reagent Compartment?

a)

50

b)

60

c)

70

d)

80

3.

How many sample positions in the STAT Table?

a)

18

b)

22

c)

24

d)

26

4.

What is the suitable temperature for reagent compartment in DxC 700 AU?

a)

2 - 4°C

b)

4 - 10°C

c)

4 - 12°C

d)

8 – 12 °C

5.

What is the cause of this flag “F”?

a)

Linearity Error in Rate Method

b)

Result is higher than the dynamic range

c)

Result is lower than the dynamic range

d)

Test pending or not analyzed

6.

Which of the following stability period for the ISE Reagents once open?

a)

ISE Reference – 2 months / ISE Buffer – 1 month / ISE Mid Standard – 2 months

b)

ISE Reference – 2 months / ISE Buffer – 2 months / ISE Mid Standard – 1 months

c)

ISE Reference – 3 months / ISE Buffer –2 months / ISE Mid Standard – 1 month

d)

ISE Reference – 1 months / ISE Buffer – 2 months / ISE Mid Standard – 1 month

7.

All the following considered as daily maintenance, except?

a)

Inspect the syringes for leaks

b)

Inspect the wash solution roller pump for leaks

c)

Clean the pre-dilution bottles

d)

Inspect and clean sample probe, reagent probes and mix bars

8.

These conditions require for re-calibrate except?

a)

Change detergent

b)

Change Calibrator

c)

The QC is outside SD limit

d)

Change reagent

9.

What is solution used for W1?

a)

2.0% Cleaning Solution

b)

0.5 M HCl

c)

2.0 % detergent/Wash Solution

d)

1.0 M HCl

10.

You can load multiple bottles of the same reagent on the analyzer. The AU identifies each bottle as a sequence. Up to how many bottle sequences of the same reagent can be on the analyzer at one time?

a)

4

b)

5

c)

6

d)

No Limit

11.

How many samples can continuously run on rack holder for DxC 700 AU?

a)

150 samples

b)

100 samples

c)

80 samples

d)

200 samples

12.

What are the important items need to be changed every time when performing W2?

a)

Reagent Probe and Sample Probe

b)

Sample Syringe and Reagent Syringe

c)

ISE solution and Detergent

d)

Pre Diluent Bottle, W2 bottle and DET -1 position. (3 bottles)

13.

How many cuvettes are there in DxC 700 AU Chemistry Analyzer?

a)

88

b)

150

c)

165

d)

170

14.

Which page is used to view sample’s results and to print the results?

a)

Data Monitor

b)

Sample manager

c)

Sample status

d)

Rack requisition

15.

What is the maximum average photocal reading (340nm) that user needs to change photo lamp?

a)

0.1

b)

0.5

c)

1.5

d)

0.9

16.

How many slots available in the STAT table?

a)

22 slots for sample, 4 for ISE, 1 for Maintenance (Clean), 2 for Reagent Blank (RB)

b)

22 slots for sample, 4 for ISE, 2 for Maintenance (Clean) ,2 for Reagent Blank (RB

c)

23 slots for sample, 6 for ISE, 3 for Maintenance (Clean), 1 for Reagent Blank (RB)

d)

23 slots for sample, 6 for ISE, 3 for Maintenance (Clean), 3 for Reagent Blank (RB)

17.

What is the correct sequence to perform an Emergency Stop and Restarting system?

a)

Turn Off CPU – Push Emergency Stop Button – Turn Off Monitor – Push Power ON Button

b)

Push Emergency Stop Button – Push Reset Button – Turn Off CPU – Push Power ON Button

c)

Push Emergency Stop Button – Turn off CPU – Push Reset Button – Push Power ON Button

d)

Turn Off CPU – Push Emergency Stop Button – Push Reset Button – Push Power ON Button

18.

What is the correct procedure to change the ISE reagent during operation (Measure mode) from reagent management screen?

a)

Click "Replace ISE Reagent" > "Yes" > Replace the reagent bottle > Click "Ok" > Perform Prime (Buffer/Mid Ref)

b)

Click "Replace ISE Reagent" > "Yes" > Scan the barcode label on the bottle> Replace the reagent bottle > Click "Ok" > Perform Prime (Buffer/Mid Ref)

c)

Click "Replace ISE Reagent" > "Yes" > Scan the barcode label on the bottle> Replace the reagent bottle > Click "Ok" > Perform ISE Calibration

d)

Click "Replace ISE Reagent" > "Yes" > Scan the barcode label on the bottle> Replace the reagent bottle > Click "Ok"

19.

Which of the following are the factors that affect a calibration result? (Multiple answer)

a)

Reagent Stability

b)

Calibrator preparation

c)

Quality control preparation

d)

Parameter Set up

20.

Which of the following are the factors that affect a Quality Control result? (multiple answer)

a)

Parameter

b)

Calibration

c)

Index

d)

Reagent

21.

What are solutions used for W2? (multiple answer)

a)

0.5% Cleaning Solution (1:10)

b)

1.0% Cleaning Solution (1:5)

c)

1 M HCl

d)

1 M NaCl

22.

What is the main function of this area?

a)

Run STAT Sample

b)

Reagent Storage

c)

Measurement Sample

d)

DI Water Tank

23.

Name the part for 1 and 2

a)

1: Reagent Probe

2 : Wash Well

b)

1: Wash Well

2: Sample Probe

c)

1: Mix Bar

2: Sample Probe

d)

1: Sample Probe

2: Wash Well

24.

Name above component

a)

Mixer/Mix Bar

b)

Reagent Probe

c)

STAT Table

d)

Reagent Syringe

25.

What is the temperature inside above component?

a)

37 °C

b)

32 °C

c)

8 °C

d)

25 °C

26.

Which one is the Wash Syringe

a)

1

b)

2

c)

3

d)

5

27.

Which tank we need to open to add the wash solution?

a)

1

b)

2

c)

3

d)

4

28.

Please name the above component

a)

Wash Solution Roller Pump

b)

Wash Syringe

c)

ISE Tubing

d)

Mixer Tubing

29.

What is the suitable action if the result have "F" flag?

a)

Dilute the sample and rerun

b)

Prepare fresh reagent

c)

Perform Calibration and QC

d)

Change ISE Reagent

30.

What is this "#" Flag mean?

a)

Insufficient sample detected

b)

Clot detected

c)

Result is lower than the dynamic range

d)

Linearity Error in Rate Method