Total questions: 104
Worksheet time: 52mins
Cylinders in storage shall be prevented from reaching temperatures of ___⁰F?
100
150
125
120
ASTM B__ Standard Specification for seamless copper water tube, copper tube (type K, L, or M)?
99
819
219
88
Standard for the storage, use and handling of compressed gases and Cryogenic Fluids in portable and stationary containers, cylinders and tanks bulk oxygen systems: NFPA__?
99
75
40
55
Vacuum receivers shall be capable of withstanding __ psi?
80
100
60
120
NFPA __ national electric code?
55
80
70
99
Standard nitrogen or instrument air at ___ - ___ psi?
50 - 60
160 - 180
150 - 180
160 - 185
The scale range of vacuum indicators shall be __"?
5-29
0-30
0-29
5-30
Valve sample port shall be provided near the oxygen concentration monitor sensor connection for sampling of the gas from the oxygen concentrator unit?
3/4"
1/4"
1/2"
3/8"
Medical air filter 98% efficient at __ or greater?
10microns
0.01microns
1micron
0.1microns
Direct and unobstructed visual inspection of the interconnecting shaft through vent and inspection openings no smaller than ___ shaft diameter(s) in size (reciprocating compressor)?
2
2.5
1.5
5
Normal vacuum gauge display ______?
12"hgv
20"hgv
5"hgv
30"hgv
Medical air intake shall be ___ above ground level?
6'
12'
20'
25'
Standing pressure test that will assure a pressure decay of less than .5% in __hrs?
12
24
8
36
Standing pressure test for med. gas is ___ above operating pressure?
20%
10%
50%
25%
Factory-installed copper tube on station outlets extending no further than 8" from the body of the terminal shall be not less than __?
1/4"
3/8"
1/2"
3/4"
Instrument air is filtered to __ microns?
0.01
0.1
1
10
Size of piping runouts to alarm panels and connecting tubing for gauges and alarm devices?
1/4"
3/8"
1/2"
3/4"
Natural vent shall be installed within __ from floor and also from ceiling?
6"
1'
4"
2'
Minimum free area of a natural vent - __in²?
120
144
72
81
ASTM B____ Standard Specification for seamless copper tubing for air conditioning and refrigeration field service, copper ACR tube?
280
88
99
819
Brazer must braze qualified procedure every _ month(s) to remain qualified?
2
12
6
18
Factory-installed copper tube on station inlets extending no further than 8" from the body of the terminal shall be not less than __?
1/4"
3/8"
1/2"
3/4"
Equal headers have not less than _ cylinders per header (manifold for gas cylinders)?
3
5
6
2
There shall be __ of clearance around storage containers, vaporizers, cabinets of a bulk site and the EOSC?
3'
10'
1'
5'
Standing vacuum test is a minimum of __" HGV to full vacuum?
20
12
5
10
Medical gas shall have less than _____ gaseous hydrocarbons?
100ppm
25ppm
200ppm
10ppm
Instrument air rated for minimum __% efficiency at 0.01 micron?
99
81
89
98
Medical gas operating pressure above ___psi is not covered by NFPA 99 code?
300
185
200
150
Liquid oxygen will expand ___ times?
2500
1700
960
860
ASSE standard for inspectors?
6020
6010
6050
6030
Bulk nitrous oxide system with a storage capacity of more than _____ft³?
3200
28,000
20,000
2800
___-___ psi: station outlet shall be designed so as to prevent the removal of the adapter until the pressure has been relieved to prevent the adapter from injuring the user or others when removed from the outlet?
185-200
200-300
175-185
150-200
Outlets in excess of _____ shall be D.I.S.S. connectors?
120
55
80
75
Medical air compressor minimum run time for inspection with a demand of approximately 25% of the rate compressor capacity shall be created to cause the compressors to cycle on and off continuously and the dryers to operate?
2
1
6
12
Minimum time for a standing pressure test for verifiers?
1hr
10mins
24hrs
30mins
Med Gas type K sized __ shall be used over 185psi?
1.5"
3/4"
3"
2"
On dissimilar metals sized __" and smaller, flux-coated brazing ribs shall be permitted to be used in lieu of applying flux to the surfaces being joined?
1/4"
3/8"
1/2"
3/4"
Medical gas main and branches shall not be less than __?
1/4"
3/8"
1/2"
3/4
Initial pressure test: pressure for gasses and vacuum piping shall be at ___ times working pressure and not less than ___ psi?
2, 150
2, 200
1.5, 150
1.5, 200
One unobstructed mechanical exhaust shall be provided within _' of floor and adjacent to the cylinders or containers?
6"
2'
3'
1'
Carbon monoxide alarm activates when CO exceeds ___?
10ppm
200ppm
100ppm
1ppm
Mains and branches in medical-surgical vacuum systems shall be not less than __?
1/4"
3/8"
1/2"
3/4"
There shall be a minimum ___ of backfill where there is no potential for damage from surface loads or surface conditions?
12"
18"
24"
36"
Flame spread index of __ or less for a CMT?
10
25
50
200
____ __ is for bulk gas, compressed gases, and Cryogenic Fluids Code?
NFPA 99
CGA 21
NFPA 55
CGA 40
Instrument air should be dry to a point of ___?
32⁰F
-40⁰F
22⁰F
-22⁰F
Nitrogen pressure for cross connection and vacuum test?
100psi
50psi
120psi
85psi
Vacuum receivers shall be capable of withstanding a __" gauge HGV test?
30
20
29
12
Instrument air dew point alarms when the temperature exceeds ___F indicating a high dew point?
32⁰
-20⁰
-22⁰
22⁰
Automatic alarms shall restart after a power loss of no more than __ seconds?
1
5
10
15
Flame spread rating of not greater than ___?
25
75
100
200
Filter shall accrue no more than __ of matter from any outlet tested at verification?
0.1mg
1mg
0.01mg
10mg
For WAGD system under __, use any noncorroding tube or ductwork be joined using any method that will result in a leak-free network when tested?
5"HGV
12"HGV
29"HGV
1"HGV
Bulk oxygen or inert gas system with a storage capacity of more than ______ft³?
1800
3200
20,000
28,000
Medical air liquid ring compressor standby header shall have attached cylinders sufficient for ____ of normal operation?
1hr
12hrs
24hrs
48hrs
Joints shall be brazed within ___ after the surfaces are cleaned for brazing?
8hrs
4hrs
30mins
12hrs
Alternate air source should be available __?
Within 1 millisecond
For up to 24hrs
24/7
Within 10 milliseconds
Instrument air standby header shall be sufficient for __hr(s) of normal operation?
3
24
1
12
(Manifold for cryogenic containers) There shall be a minimum of _ cylinders on the reserve header?
2
4
3
6
A system with nonstandard pressures shall be tested at ___ higher or lower than other system?
50%
20%
10psi
20psi
Med air intake shall be ___ away from an exhaust or vent?
25'
10'
15'
20'
Cancelable audible alarm with minimum level of __dba at 3'?
80
85
105
120
Standing pressure test for med gas and vacuum systems is for __?
1hr
24hrs
6hrs
2hrs
Test coupons shall be welded and inspected, as a minimum, at the start of work and every _hr(s) thereafter, our when the machine is idle for more than 30 minutes, and at the end of the work period?
6
1
3
4
Labels shall be a minimum of __ tall on liquid containers?
3"
1.5"
1"
2"
Minimum pressure rating for memory metal and axially swaged fittings?
200psi
185psi
300psi
150psi
Standard Specification for seamless and welded austenitic stainless steel tubing for general service?
304
306
304, 306
304, 316
Temperature ratings on memory metal and axially swaged fittings?
860⁰F
1200⁰F
1160⁰F
1000⁰F
Bulk nitrous oxide system with storage capacity of more than _____ lbs?
1800
20,000
3600
3200
ASSE standard for maintenance personnel?
6050
6040
6030
6020
Temperature of liquid oxygen?
-22⁰F
-35⁰F
-225⁰F
-300⁰F
Maximum voltage rating for areas having pipe installed?
440
800
600
15kv
ASTM B____ Standard Specification for seamless copper tube for medical gas systems, copper medical gas tubing (type K, L)?
55
819
88
219
Sch. __ PVC is the minimum for vacuum systems in a Category 3 facility?
20
40
80
120
___ mean(s) of entry/exit(s) at an outdoor central supply system site greater than 200ft²?
Two
Three
One
Four
Air intakes shall be located a minimum of ___ above roof, door, window, exhaust, or another intake?
3'
25'
10'
15'
Dew points shall be monitored and shall activate a local alarm and all master alarms when the dew point at system delivery pressure exceeds __⁰F?
-22
35
20
-44
Standard air pressure should be at __ - __ psi?
60-80
55-60
70-85
50-55
Central supply systems for CO² using permanently installed containers with product capacities greater than ____lbs is a bulk system?
1000
3200
3600
20,000
Alarm indication when the medical surgical vacuum pressure in the main line of each vacuum system drops to or below ___HGV?
5"
20"
12"
10"
WAGD pipe joined to medical surgical vacuum shall be connected a minimum distance of __ from any vacuum inlet?
5'
10'
15'
20'
Medical air has ___ - ___ % oxygen?
17-25
19.5-32.5
21.5-25.6
19.5-23.5
Medical gas tube type K shall be used on pressures greater than ___ psi?
160
185
200
175
ASSE standard for instructors?
6010
6040
6030
6050
___% oxygen is oxygen enriched?
25.6
22
35
23.5
Relief valves for medical gas systems with less than ____ft³ med gas shall be permitted to be diffused locally by means than will not restrict flow?
1200
3200
20,000
3000
Pipe labels shall be installed at intervals of not more than __'?
20
10
15
25
Surfaces exhibiting a flame spread index not greater than __ is limited combustible?
20
25
50
200
Flexible hose minimum burst gauge pressure?
860psi
1000psi
1160psi
1200psi
ASSE standard for installers?
6030
6050
6010
6020
No plastic or polymeric is allowed in systems with ___psi or greater?
185
300
200
350
Area alarms will indicate when an increase or decrease of __% of total psi occurs?
20
15
25
10
Brazed joints shall be made using a brazing alloy that exhibits a melting temperature in excess of _____F?
1200⁰
1160⁰
1000⁰
890⁰
ASSE standard for verifiers?
6010
6040
6020
6030
There shall be a __" minimum backfill over piping outside of a building?
12
36
18
42
Label wraparound pipe ___ degrees on liquid containers?
-300
212
185
360
Nitrous oxide & CO² cylinders in storage shall be prevented from reaching lower than ___F?
-20
20⁰
35⁰
-40
Vacuum receivers shall be capable of withstanding a gauge pressure of ___psi & a vacuum of ___"HgV
60, 30
30, 60
150, 24
300,12
EOSC size for ___% of system demand?
100
99
85
75
Relief valve setting shall be ___ above normal operating pressure?
75
50
60
20
MGR ankles shall be made of material with a melting point of at least ____⁰F?
890
1000
1200
2100
If the medical-surgical vacuum source produces WAGD, the total concentration of oxidizers (Oxygen and Nitrous Oxide) shall be maintained below ___
19.5%
23.5%
23.6%
30%
Efficiency of medical air filters at 1 micron or greater?
99%
89%
75%
98%