Total questions: 78
Worksheet time: 46mins
What is this a physical description of?
-Consists of a case, power transformer, power supply, temperature control module, and an internal sampling system. The case is made of cast aluminum with a gray epoxy finish. Inlet and outlet gas connections are provided for sample gas flow through the analyzer.
ORSAT Test Set
Infrared Spectrophotometer
Oxygen/Nitrogen Analyzer (Servomex)
FCS 2001
What is this physical description of?
-O2 and N2 chemicals
-Glassware and holders
-Tubing (for various connections)
(Cabinet)
ORSAT Test Set
Oxygen/Nitrogen Analyzer
FCS 2001
Infrared Spectrophotometer
What is this a physical description of?
-Differs depending on model.
(Sorry, very vague, that's all it said)
Oxygen/Nitrogen Analyzer
ORSAT Test Set
Infrared Spectrophotometer
FCS 2001
What is this a physical description of?
-Bottle: DOT rated st 1800 PSIG
-Sample extraction valve
-Clean out plug
-Pressure gauge
-Coupling fitting
-Burst Disk
-Fill hose and coupling
-Spare kit w/ seal rings
-Case assembly
-Purge fitting
Oxygen/Nitrogen Analyzer
ORSAT Test Set
Infrared Spectrophotometer
FCS 2001
What is this the functional operation of?
-Designed to continuously monitor the contents of a sample gas stem. Accepts a gas sample, and then produces an output current which is proportional to the magnetic susceptibility of the sample.
ORSAT Test Set
Oxygen/Nitrogen Analyzer (Servomex)
Infrared Spectrophotometer
FCS 2001
What is this the functional operation of?
-Used to test gaseous O2 or N2 for purity. In either gas test, oxygen is selectively removed from the gas stream through a chemical reaction. Measurement of the remaining gas after reaction provides purity level of the gas sample.
ORSAT Test Set
Oxygen/Nitrogen Analyzer (Servomex)
Infrared Spectrophotometer
FCS 2001
What is this the functional operation of?
-Capable of measuring contaminants and their parts per million level in a gas sample. Maintained and operated by air wing personnel on aircraft carriers.
Oxygen/Nitrogen Analyzer (Servomex)
ORSAT Test Set
Infrared Spectrophotometer
FCS 2001
What is this the functional operation of?
-Charging manifold consists of two cups, one inside the other, inner cup is a high pressure vessel. Liquid being sampled is allowed to flow between the two cups until the sampling cup is cooled. Inlet valve is opened to fill the inner cup, and then closed to retain a known volume of liquid in the cup.
Oxygen/Nitrogen Analyzer (Servomex)
ORSAT Test Set
Infrared Spectrophotometer
FCS 2001
What is this term?
-The ___ ___ of ABO requires frequent testing and continuous surveillance.
Procurement Limits
Quality Control
Contamination Control
3M Management
What is this term?
-Having a knowledge and understanding of how contaminants are concentrated in LOX and the sources of increasing contamination is essential to the effective control of contamination.
Quality Control
3M Management
Contaminant
Contamination Control
What is this term?
-Undesirable foreign mattering a gas or a liquid.
Contaminant
Purity
Limit
FOD
What is this term?
-Clean up of the process air stream, usually accomplished by refrigeration and physical adsorption.
Contamination
Purification
Vaporization
Procurement
What kind of purification is this?
-A simple way of removing moisture and carbon dioxide by condensation of moisture and freezing of CO2 in either regenerators or reversing heat exchangers.
Refrigeration Purification
Physical Adsorption
What kind of purification is this?
-Where a layer of gas one or more molecules thick is attracted to a solid surface and held there by molecular forces. This is useful in the cleanup of gaseous impurities at or near their condensation temperature.
Refrigeration Purification
Physical Adsorption
What is this term?
-The increase of contaminants due to LOX vaporization cannot be prevented in the storage tank. However the rate of increasing contamination from this source shall be controlled by minimizing heat additions due to handling and transfer through proficient and careful performance of operations.
Condensation Contamination
Procurement Contamination
Vaporization Contamination
Convention Contamination
What is this term?
-The addition of contaminants from the environment shall be prevented by careful and proficient handling and transfer operations.
Procurement Limits
Use Limits
Emergency Use Limits
Handling and Transfer
What is this term?
-Purity and contamination requirements for liquid and gaseous ABO, whether it is generated by Navy and Marine Corps activities or procured form commercial sources.
Handling and Transfer
Procurement Limits
Use Limits
Emergency Use Limits
What is this term?
-Applicable to storage tanks, servicing equipment and converters.
Handling and Transfer
Procurement Limits
Use Limits
Emergency Use Limits
What is this term?
-Concerning use of ABO under emergency conditions when liquid ABO contains a greater concentration of contaminants than the use limit or has a low purity.
Handling and Transfer
Procurement Limits
Use Limits
Emergency Use Limits
What is this reference?
-Covers the minimum DOD procurement limits and requirements for two types of Aviators Breathing Oxygen or ABO.
(Type 1: Gaseous, Type 2: Liquid)
MIL-PRF-27210
NAV AIR-A6-332AO-GYD-000
NSTM CH. 550
FED. SPEC. A-A-59155
What is this reference?
-Provides information, guidance, instructions and procedures for quality control of types 1 and 2 ABO.
-Contains the minimum procurements and use limits as prescribed by the MIL-PRF-27210. Contains expanded limits of LOX servicing carts and LOX converters for type 2 ABO.
-Sets guidelines for emergency use limits for type 1 and 2 ABO.
MIL-PRF-27210
NAV AIR-A6-332AO-GYD-000
NSTM CH. 550
FED. SPEC. A-A-59155
What is this reference?
-Contains requirements, procedures and frequencies for the sampling and testing of types 1 and 2 ABO.
-Must meet the minimum guidelines as prescribes in the MIL-PRF-27210, with the exception of methane.
MIL-PRF-27210
NAV AIR-A6-332AO-GYD-000
NSTM CH. 550
FED. SPEC. A-A-59155
What is this reference?
-Provides the federal specifications for nitrogen.
-Type 1: Gaseous
Type 2: Liquid
-Grade A: 99.95% Pure
-Grade B: 99.50 Pure
-Grade C (Type 1 only): 99.50% Pure, moisture content not specified
-Class 1: Oil Free
-Class 2: Oil Tolerant
MIL-PRF-27210
NAVAIR-A6-332AO-GYD-000
NSTM CH. 550
FED. SPEC. A-A-59155
What kind of generating plant is this?
-ABO procured must meet the requirements of the current issue of the MIL-PRF-27210.
Commercial
Navy and Marine Corps
What kind of generating plant is this?
-ABO produced must meet the requirements cited in the NAVAIR A6-332AO-GYD-000.
-The only exception is methane, which is cover in the NSTM CH. 550.
Commercial
Navy and Marine Corps
What kind of testing requirement is this?
-Analyzed during production. Shall be accomplished using either an electronic bulk ___ analyzer or an ORSAT chemical test. Shall be tested prior to beginning O2N2 transfer to bulk storage.
Purity
Acetylene
Odor Test
Analysis for Trace Contaminants
What is this testing requirement?
-Generating plants will be tested for ___ during O2 and N2 production. Within every 4 hours after starting production, and every 24 hours thereafter.
Purity
Acetylene
Odor Test
Analysis for Trace Contaminants
What kind of test requirement is this?
-If the producer has a liquid sample valve, it shall be tested for ___ within 4-8 hours after start of production. If it does not have liquid sample valve, the sample will be taken at the tank.
Purity
Acetylene
Odor
Analysis for Trace Contaminants
What is this testing requirements?
-If the producer has a liquid sample valve, the sample will be taken from the producer and product shall be tested for contaminants witting 4-8 hours after start of O2 production.
Purity
Acetylene
Odor Test
Analysis for Trace Contaminants
What procurement limit is this?
Type 1: 10.0p/m
Type 2: 5.0p/m
(a)
What procurement limit is this?
Type 1: 50.0p/m
Type 2: 25.0p/m
(a)
What procurement limit is this?
Type 1: 0.1p/m
Type 2: 0.05p/m
(a)
What procurement limit is this?
Type 1: 0.4p/m
Type 2: 0.2p/m
(a)
What is this procurement limit?
Type 1: 6.0p/m
Type 2: 3.0p/m
(a)
What procurement limit is this?
Type 1: 4.0p/m
Type 2: 2.0p/m
(a)
What procurement limit is this?
Type 1: 2.0p/m
Type 2: 1.0p/m
Refrigerants
Other/Solvents
What procurement limit is this?
Type 1: 0.2p/m
Type 2: 0.1p/m
Refrigerants
Solvents/Other
How often will an idle tank that contains liquid O2 be tested for acetylene?
Hourly
Daily
Weekly
Bi-Weekly
True/False
A particulate test is only required when contamination is suspected.
True
False
What use limit is this?
Type 1: 20.0p/m
Type 2: 10.0p/m
(a)
What use limit is this?
Type 1: 100.0p/m
Type 2: 50.0p/m
(a)
What use limit is this?
Type 1: 0.2p/m
Type 2: 0.1p/m
(a)
What use limit is this?
Type 1: 0.8p/m
Type 2: 0.4p/m
(a)
What use limit is this?
Type 1: 10.0p/m
Type 2: 6.0p/m
(a)
What use limit is this?
Type 1: 6.0p/m
Type 2: 4.0p/m
(a)
What use limit is this?
Type 1: 4.0p/m
Type 2: 2.0p/m
Refrigerants
Solvents/Other
What use limit is this?
Type 1: 0.4p/m
Type 2: 0.2p/m
Refrigerants
Solvents/Other
True/False
All gaseous O2/N2 cylinders do not need to be tested for odor prior to filling.
True
False
What is this temperature operating range?
14°F min-104°F max
Sample Temp
Ambient Temp
Storage Temp
Internal Case Temp
What is this temperature operating range?
14°F min-122°F max
Sample Temp
Ambient Temp
Storage Temp
Internal Case Temp
What is this temperature operating range?
-40°F min-185°F max
Sample Temp
Ambient Temp
Storage Temp
Internal Case Temp
What is this temperature operating range?
122°F (automatically controlled by power supple and temp control)
Sample Temp
Ambient temp
Storage Temp
Internal Case Temp
What are the power requirements of an oxygen analyzer?
115 VAC
110 VAC
48-62 Hz
300 A
250 A
What is this fluid flow requirement?
50-150 cc/m
Sample Flow
Bypass Flow
Max Flow
What is this fluid flow requirement?
400-7900 cc/m
Sample Flow
Bypass Flow
Max Flow
What is this fluid flow requirement?
1000 cc/m
Sample Flow
Bypass Flow
Max Flow
What is this pressure operating range?
0.2 psig min
1-2 psig reccomended
20 psig max
Sample Inlet
Sample Vent
What is this pressure operating range?
12-18 psia
Sample Inlet
Sample Vent
NAME. THIS. COMPONENT.
Physical Appearance: Rectangular, gray epoxy finish, two doors (large door electrical, small door controls)
Location: Usually next to/near control panel/gauge board, ideal installation side is vibration free where fluctuation of ambient temperature is minimal
Construction Features: Drip proof, corrosion resistant, cast aluminum, approximately 13" high x 18" wide x 10" deep and 60 lbs
Function: Provides a mounting plate and protection from moisture and temperature fluctuation
Heat Indicator
Bypass Flow Meter
Case
Piping
NAME. THIS. COMPONENT.
Physical Appearance: Graduated glass tube with large ball float
Location: Front, upper, right corner
Construction Features: Graduated from 1-5, corresponding 1-7 liters/minute
Function: Indicates ___ flow rate
Bypass Flow Meter
Gain Check Switch
Meter Test Switch
Sample Flow Meter
NAME. THIS. COMPONENT.
Physical Appearance: Graduated glass tube with small ball float
Location: Front, low center
Construction Features: Graduated from 1-5, corresponding to 50-150 cc/minute
Function: Indicates ___ flow rate
Bypass Flow Meter
Case
Meter Test Switch
Sample Flow Meter
NAME. THIS. COMPONENT.
Physical Appearance: Green LED on black background
Location: Inside the case in the bottom left window
Construction Features: Green LED
Function: Indicated when heaters on, light will flash when interior casing stabilizes at 122°F/50°C, stabilization will take approximately 4 hours
Case
Sample Flow Meter
Heat Indicator
Internal Temperature Set Point Control
NAME. THIS. COMPONENT.
Physical Appearance: Black, circular pushbutton
Location: Behind locked access door on the front panel
Construction Features: Momentary contact switch
Function: Quick check for analyzer servo loop gain, when actuated and error signal is introduced into the loop, should not change the output by more than 1.5% oxygen
Heat Indicator
Safety Stop
Meter Test Switch
Gain Check Switch
NAME. THIS. COMPONENT.
Physical Appearance: Black, circular pushbutton
Location: Behind the locked access door on the front panel
Construction Features: Momentary contact switch
Function: When actuated, digital meter should read +188.88
Meter Test Switch
Bypass Flow Meter
Gain Check Switch
Sample Flow Meter
NAME. THIS. CONTROL.
Physical Appearance: Red, circular knobs (2)
Location: Behind the locked access door on the front panel
Construction Features: Potentiometer, a 3 terminal resistor with 1 or more sliding center contacts that function as adjustable voltage dividers
Function: Calibration span, calibration zero
Internal Temperature Set Point Control
Calibration Span and Calibration Zero Controls
Output Range Selector Switch
Trim-Pot A and Trim-Pot B
NAME. THIS. COMPONENT.
Physical Appearance: Stem and slotted adjustment screw
Location: Behind the locked access door on the front panel
Construction Features: Potentiometer with a slotted stem
Function: Used to change the factory-set temperature inside the oxygen analyzer module, adjust counter-clockwise to lower temperature, adjust clockwise to raise temperature
Calibration Span and Calibration Zero Controls
Output Range Selector Switch
O2N2 Selector Switch
Internal Temperature Set Point Control
NAME. THIS. COMPONENT.
Physical Appearance: Rotating black knob with 7 settings on a scale
Location: Behind the locked access door on the front panel
Construction Features: Resistor with 7 contact points
Function: Selects the range of O2 that will be proportional to the voltage output of the 510 module as indicated on the digital readout meter
Output Range Selector Switch
Internal Temperature Set Point Control
Trim-Pot A and Trim-Pot B
O2N2 Selector Switch
NAME. THIS. COMPONENT.
Physical Appearance: Toggle switch
Location: Behind the locked access door on the front panel
Construction Features: 2 points
Function: Selects the mode of analyzer operation and changes the voltage output
Internal Temperature Set Point Control
O2N2 Selector Switch
Trim-Pot A and Trim-Pot B
Output Range Selector
NAME. THIS. COMPONENT.
Physical Appearance: Slotted adjustment screw
Location: Behind locked access door on the front panel
Construction Features: Adjustment screw electrically connected to variable resistors
Function: One is the Zero adjustment and one is the Span adjustment
Output Range Selector Switch
Calibration Span and Calibration Zero Controls
Internal Temperature Set Point Control
Trim-Pot A and Trim-Pot B
Which piping is this?
Physical Appearance: 1/8" NPT stainless steel
Location: Right Side of analyzer
Construction Features: Only non magnetic piping to prevent damage to analyzer, should be suitable for use with pure oxygen
Sample
Vent
Which piping is this?
Physical Appearance: 1/4" NPT stainless steel
Location: Bottom, right side of analyzer
Construction Features: Only non magnetic piping to prevent damage to analyzer, should be suitable for use with pure oxygen
Sample
Vent
What kind of valve is this?
Physical Appearance: Silver cap labeled "B"
Location:Analyzer front, lower right corner
Construction Features: Needle valve
Function: Controls the flow of gas to the analyzer cell
Bypass
Sample
What kind of valve is this?
Physical Appearance: Silver cap labeled "S"
Location:Analyzer front, lower right corner
Construction Features: Needle valve
Function: Regulates the flow of gas to the analyzer measuring cell
Bypass
Sample
NAME. THIS. COMPONENT.
Physical Appearance: Red LED display on a black background
Location: Analyzer front, lower left behind the window
Construction Features: Seven character LED display
Function: Reads the current output, as processed through the analyzer circuit and displays it as a percentage to 1/100%
Meters/Digital Readout
Analyzer Assembly
Output Range Selector Switch
Internal Set Point Control
NAME. THIS. COMPONENT.
Physical Appearance: Rectangular box with adjustment knobs on the left side
Location: Upper right, inside main casing, access through front panel
Construction Features: 2 nitrogen filled quartz spheres arranged in the form of a dumbbell with a single turn of platinum wire in place around the dumbbell, a rigid taut band of platinum ribbon, attached at mid-point, forming a suspension system. Permanent magnet with specially shaped pole pieces, a mirror, 2 photocells, a light source and an amplifier
Function: Oxygen molecules are attracted to the strongest part of the magnetic field around the dumbbell.
Meters/Digital Readout
Output Range Selector Switch
O2N2 Selector Switch
Analyzer Assembly
What are the power sources of an analyzer interface?
115 VAC
110VAC
48-62 Hz
300V Amps
330V Amps
True/False
Calibration gas purity is normally 99.9997%
True
False
How often, at a minimum, is the purity reading of the analyzer recorded?
A minute
Every hour
Every 2 hours
Every 4 hours