1.

Instrument air systems shall be permitted to interconnect with medical air systems

provided the system has ______.

a)

separate source valves for each system

b)

interconnection is prohibited

c)

been approved by the manufacturer

d)

at least two or more compressors

2.

The scale range of positive pressure analog indicators shall have the normal reading fall within the ______ percentage of the gauge scale.

a)

1/4

b)

1/3

c)

1/2

d)

2/3

3.

What does it mean when equipment, material, or services are shown as listed?

a)

They are identified and published by an organization that is acceptable to the AHJ

b)

They have been found tolerable for the purpose described in the manufacturer’s specifications

c)

They have been tested and approved by the construction supervisor

d)

They have been tested by a medical gas verifier

4.

Fittings, valves, and other components for oxygen use in a Category 1 distribution system shall be __________ until prepared for installation.

a)

delivered sealed, labeled, and kept sealed

b)

kept at a minimum inventory level

c)

stored in a locked room or secure facility

d)

plugged or capped by the manufacturer

5.

Installers of Category 1 medical gas piping systems must be qualified to ASSE 6010. What is the certification requirement for installers of a category 3 system?

a)

ASSE 6010

b)

ASSE 6020

c)

ASSE 6030

d)

Category 3 does not require certification

6.

Metallic flexible joints are being used in the pipeline for vibration control.

Which of the following is required of such joints?

a)

Received on the jobsite with certification of cleanliness

b)

All that are listed are required

c)

Provided with extensions for brazing

d)

Able to withstand 1000° F

7.

NFPA 99 specifies that the medical gas piping system shall be protected against all EXCEPT which of the following?

a)

Physical damage

b)

Freezing

c)

Combustible liquids

d)

Corrosion

8.

A signal from each ________ system shall be connected to the master alarm panels to indicate a problem is present with the source equipment.

a)

area alarm

b)

pressure switch

c)

local alarm

d)

anesthetizing

9.

Where existing, the portion of the supply equipment that automatically supplies the system when the primary and secondary are unable.

a)

Back-up source

b)

Secondary source

c)

Reserve supply

d)

Emergency supply

10.

Which of these operations would NOT breach the vacuum system?

a)

Addition of a system component

b)

Removal of a system component

c)

Replacement of a system component with an identical one

d)

Testing a system component with a leak detecting solution

11.

Liquid ring compressors, service water and seal water shall be treated to control _________.

a)

waterborne pathogens

b)

chlorine

c)

hyper chlorination

d)

all that are listed

12.

Cylinder and cryogenic liquid container outlet connections shall be affixed in such a manner as to be integral to the valve, un-removable with ordinary tools or so designed as to render the attachment __________ when the removal of the outlet connection is attempted.

a)

universal

b)

unusable

c)

reusable

d)

unsuitable

13.

The medical air intake shall be located a minimum of ________ feet from doors, windows or other openings in the building.

a)

10

b)

15

c)

20

d)

25

14.

Heating, cooling, ventilating and process systems serving spaces or providing healthcare functions shall be provided in accordance with _____________.

a)

ASHRAE 170

b)

ASSE 6000

c)

NFPA 55

d)

NFPA 99

15.

New medical gas and vacuum pipeline distribution systems shall not be connected to the existing in-use systems, until which installer performed tests have been completed?

a)

All

b)

No required tests

c)

Initial pressure

d)

Standing pressure

16.

New or replacement shutoff valves for medical gas pressure piping systems shall _________.

a)

be quarter turn valves

b)

be of brass or bronze construction

c)

have extensions for brazing

d)

all that are listed

17.

The auxiliary source connection shall be _________ and secured.

a)

permanently closed

b)

normally closed

c)

normally open

d)

none that are listed

18.

Vacuum systems deemed breached shall be subject to inspection and testing of the specific area and the area located ________ of the point of intrusion

a)

downstream

b)

near the surrounding area

c)

upstream

d)

all that are listed

19.

Central supply systems for nitrous oxide and carbon dioxide using cylinders, shall be prevented from reaching temperatures lower than _______° F.

a)

125

b)

-20

c)

20

d)

32

20.

Fittings with internal seals shall be permitted where they meet which of the following requirements?

a)

Shall be brass or bronze

b)

Shall be cleaned for oxygen service when used for patient or support gases

c)

Shall be installed only where visible and accessible, allowing for testing and servicing

d)

Shall be cleaned for oxygen service when used for patient or support gases, and shall be installed only where visible and accessible, allowing for testing and servicing

21.

Copper tube hangers for medical gas and vacuum distribution piping shall be ___________.

a)

supported from the building structure

b)

sized for copper tube

c)

permitted to be plastic coated in potentially damp locations

d)

all that are listed

22.

Where compressors that separate the oil containing section from the compression chamber are used, each compressor shall be monitored for contaminants for which of the following?

a)

Gaseous hydrocarbons

b)

High temperature

c)

Liquid hydrocarbons

d)

Gaseous hydrocarbons and high temperature

23.

Piping between the vacuum pump and the source shutoff valve shall be permitted to be ______.

a)

brass pipe

b)

galvanized steel pipe

c)

black steel pipe

d)

any of the listed materials

24.

An assembly of equipment for supplying compressed gas (consisting of, but not limited to, storage containers, pressure regulators, pressure relief devices,vaporizers, manifolds, and interconnecting piping) that terminates where the gas, at system pressure, first enters the main line. The storage containers are either stationary or movable and include unconnected reserves on hand at the site, and the source gas is stored as a compressed gas or cryogenic fluid. What type of system is this?

a)

Bulk system

b)

Compressed air system

c)

Vacuum system

d)

Medical gas system

25.

A low pressure, vacuum-insulated vessel containing gases in liquid form. What is this the definition for?

a)

Appliance

b)

Container

c)

Cylinder

d)

Double insulated appliance

26.

The medical air intake shall be located a minimum of ___ feet from all vehicular exhausts and noxious fumes.

a)

10

b)

15

c)

20

d)

25

27.

Computer systems used as a master alarm shall be continuously attended or shall provide remote signaling of responsible parties through ______________.

a)

pagers

b)

telephones

c)

cell phones

d)

any of the listed methods

28.

Liquid ring compressor systems shall have a reserve medical air standby by header (1HR) or a __________ as a backup.

a)

condensate trap

b)

compressor

c)

manifold

d)

bulk system

29.

A chemical element that, at normal atmospheric temperatures and pressures, exists as a colorless, odorless and tasteless gas, and comprises about 21% of the earth’s atmosphere by volume.

a)

Helium

b)

Oxygen

c)

Nitrous oxide

d)

Nitrogen

30.

What percent of the zones shall be tested at the outlet most remote from the source to verify the piping particulate test?

a)

100%

b)

75%

c)

50%

d)

25%

31.

In central supply systems using gas cylinders, each header shall have a ________ valve to prevent backflow and to permit service to the header.

a)

Relief

b)

Ball

c)

Bypass

d)

Check

32.

ASSE 6000 Annex J provides “Recommended Guidelines for Conducting Planned Medical Gas Systems Shutdown and Temporary Backfeed.” Annex J guidelines do NOT include a list of responsibilities for which types of worker?

a)

Verifier

b)

Medical Gas Generalist

c)

Installer

d)

Shutdown Coordinator

33.

Which source shall NOT be used to power the venturi of a WAGD system?

a)

Water

b)

Inert gas

c)

Instrument air

d)

Medical air

34.

Which NFPA code should you refer to for Life Safety?

a)

101

b)

90A

c)

170

d)

45

35.

Receivers for medical air shall be equipped with ___________.

a)

Relief valves

b)

Automatic and manual drains

c)

Sight glass

d)

All that are listed

36.

To facilitate service, testing or replacement, all pressure switches and main line gauges downstream of the source valve shall be provided with a _______________.

a)

check valve

b)

shutoff valve

c)

D.I.S.S. fitting

d)

demand check fitting

37.

Which of these is an NFPA 99 requirement for labeling of pipes in a medical gas system?

a)

Reflective label

b)

Fixed label width

c)

Label shall wrap completely around the pipe

d)

Color coding

38.

Service manuals, procedures and instructions provided by the _________ shall be used in the maintenance of equipment.

a)

Manufacturer

b)

ASSE

c)

AHJ

d)

None that are listed

39.

A tool is allowed to be used by the installer to mechanically limit the makeup depth of the fitting to be brazed. Under what conditions is this method permitted for the installation of medical gas piping?

a)

Where limited depth is not less than the minimum cup depth per ANSI/ASME B16.50

b)

Only where BCuP brazing rod is used

c)

Only on vacuum systems

d)

Only on final line connections

40.

The moisture concentration of the outlet test shall not exceed _____ ppm or an equivalent pressure dew point of -12° C (10° F) at a gauge pressure of 345 kPa (50 psi).

a)

150

b)

300

c)

500

d)

600

41.

Station outlets that operate at pressures other than standards gauge pressures listed in Table 5.1.11, shall have which additional requirements for outlet labeling?

a)

Be covered to limit access

b)

Be located away from other outlets

c)

Include the operating pressure along with the name of the gas

d)

All that are listed shall apply

42.

Those sections or portions of the piping system that serve a room or group of rooms on the same story of the facility. This is the definition for which of the following?

a)

Risers

b)

Vent lines

c)

Main lines

d)

Branch (lateral) lines

43.

In a medical gas system with emergency reserve, which alarm system(s) will activate when, or just before, the changeover to the reserve supply occurs?

a)

Local & Master

b)

Area & Master

c)

Local

d)

Master

44.

In those portions of systems intended to handle oxygen at gauge pressures greater than _______ psi, interconnecting hose shall contain no polymeric materials.

a)

350

b)

435

c)

450

d)

375

45.

The final connection between new piping and an existing system may be tested with the system gas, and must be checked with a leak detectant that does NOT contain which of these agents?

a)

Ammonia

b)

Freon

c)

Acetone

d)

Chlorine

46.

According Annex D in the ASSE 6000 sufficient spare parts should be maintained to _____.

a)

accommodate replacement of expendable parts

b)

allow for a one month supply

c)

allow for a one year supply

d)

there is no requirement to maintain spare parts

47.

Medical air dryers shall be provided with a ________ downstream of each dryer for maintenance.

a)

sample port

b)

relief port

c)

bleed off valve

d)

none of the listed parts

48.

Which is a permitted use of the medical-surgical vacuum system?

a)

Waste anesthetic gas disposal

b)

Liquid or debris disposal

c)

Non-medical or non-surgical applications

d)

Vacuum steam condensate return

49.

The medical-surgical vacuum exhaust shall be turned down and screened. The screening shall be fabricated using what type of material?

a)

Steel mesh

b)

Steel wire

c)

Any non-corroding material

d)

Any that are listed

50.

The accuracy of digital indicators shall be +/- _____% of the operational pressure at which they are used.

a)

1

b)

3

c)

5

d)

7

51.

Immediately outside each Category 1 space and anesthetizing location, there shall be located a(n) ______________ for each medical gas, vacuum and WAGD system.

a)

area alarm

b)

in-line valve

c)

relief valve

d)

none that are listed

52.

A facility is in the planning stages of installing an emergency oxygen supply connection (EOSC). How many feet is the minimum clearance around this connection?

a)

1

b)

3

c)

5

d)

1.5 x the width of the auxiliary source for service to the system

53.

Exists at cryogenic temperature, approximately -184.4° C (-300° F) at atmospheric pressure. It retains all of the properties of gaseous oxygen, but, in addition, when allowed to warm to room temperature at atmospheric pressure, it will evaporate and expand to fill a volume ______ times its liquid volume.

a)

8

b)

86

c)

100

d)

Greater than 800

54.

According to the ASSE 6000, which is likely to clog vacuum piping?

a)

Activation of reserve supply

b)

Filter changes

c)

New valve parts

d)

Dust or body fluids

55.

Medicals gas system maintenance personnel shall be required to demonstrate a basic working knowledge of all __________ listed in ASSE standard 6001, as used in NFPA 99.

a)

braze requirements

b)

dental equipment

c)

installation methods

d)

terminology

56.

Category 1 medical-surgical vacuum sources shall, at a minimum, consist of ____ or more vacuum pumps.

a)

1

b)

2

c)

3

d)

4

57.

Spare parts shall be maintained for all EXCEPT which of the following?

a)

Valves

b)

Outlets

c)

Alarms

d)

Brazed fittings

58.

The vacuum exhaust shall not discharge __________________.

a)

at a different level from air intakes

b)

outside

c)

in an area of public assembly

d)

on a roof

59.

The reserve header in a cryogenic liquid cylinder manifold system, shall have sufficient cylinder connections for an average day’s supply, but NOT fewer than how many connections?

a)

1

b)

2

c)

3

d)

4

60.

To assure that all ambient air has been removed from the pipeline prior to brazing, a(n) ___________ shall be used to verify the effectiveness of the nitrogen purge.

a)

oxygen analyzer

b)

nitrogen analyzer

c)

breathalyzer

d)

air analyzer

61.

It is the responsibility of the ________ to furnish documentation certifying that all installed piping materials are cleaned for oxygen service.

a)

authority having jurisdiction

b)

installer

c)

contractor

d)

verifier

62.

Medical gas central supply systems are NOT permitted to be located where?

a)

In the same storage facility as medical gas cylinders

b)

Remote from the medical gas system use points

c)

In an unlocked location

d)

Outdoors

63.

The __________ valve shall be the line separating the applicability between NFPA 55 and this code.

a)

Source

b)

Main

c)

In-line

d)

none that are listed

64.

A minimum of 1000 L (35 cubic feet) of gas shall be filtered through a clean, white ________ micron filter at a minimum flow rate of 100 NI/min (3.5 SCFM).

a)

0.5

b)

0.45

c)

0.55

d)

0.47

65.

Who shall witness the installer perform the standing pressure test?

a)

Installation contractor superintendent

b)

Medical gas verifier

c)

The authority having jurisdiction or its designee

d)

Medical gas inspector, or verifier, or the authority having jurisdiction or its designee

66.

Oxygen and medical air outlets serving Category 1 spaces shall allow a transient flow rate of 170 SLP (______ SCFM) for 3 seconds.

a)

3.5

b)

5

c)

6

d)

7

67.

The installation of medical gas and vacuum systems shall be made by experienced, qualified and competent installers. The installers shall meet the requirements of _____________.

a)

ASME Section IX

b)

ASSE Standard 6010

c)

AWS B31.3

d)

none of the listed standards meet the requirements

68.

Compressed gas cylinders must meet which of the following criteria, in order to be acceptable for use in an acute health care facility?

a)

Be appropriately identified by labels or stencils in accordance with CGA C-7 criteria

b)

Be constructed in accordance with specifications and regulations of OSHA

c)

Be appropriately color coded in accordance with U.S.D.A. requirements

d)

Utilize gas specific CGA connections

69.

Fittings for positive pressure medical gas systems shall be cleaned for oxygen service by _________ in accordance with CGA G-4.1.

a)

the manufacturer

b)

a supplier

c)

a third party agency other than the manufacturer

d)

all of the listed agencies are acceptable

70.

A chemical process of oxidation that occurs at a rate fast enough to produce heat and usually light in the form of either glow or flame. What is this the definition of?

a)

Combustion

b)

Combustible

c)

Friction

d)

Ignition

71.

Which of these is a requirement for area alarm panels?

a)

Located at a nurses station and in the engineering department

b)

Monitored 24/7 continuously

c)

Provide visual and audible indication

d)

None that are listed are required

72.

With the approval of the authority having jurisdiction, in what areas shall the shutoff valves be permitted to be secured to prevent inappropriate access?

a)

Psychiatric or pediatric

b)

Critical care

c)

Maternity

d)

Anesthetizing

73.

Seals necessary for the operation of the valve and prevention of leaks are _____________.

a)

replaceable

b)

removable

c)

reflexible

d)

gas specific

74.

When designing and constructing central supply systems, the storage shall __________.

a)

have a minimum of two entry/exits if outdoors

b)

secure all cylinders

c)

have a minimum 1-hour fire resistance rating

d)

all that are listed apply

75.

Regular preventative maintenance shall be scheduled to check the vacuum pressure performance of each medical-surgical vacuum station inlet. This schedule shall be as determined by the __________.

a)

AHJ

b)

Medical gas inspector

c)

Medical gas verifier

d)

Facility maintenance staff

76.

While brazing medical gas copper joints, a purge gas shall flow continuously to prevent the formation of copper oxide on the inside surface of the joint. What shall be used as a purge gas?

a)

Any inert gas

b)

Oil-free, dry Argon MD

c)

Oil-less, dry Nitrogen NF

d)

None that are listed

77.

The system designer shall size the piping such that calculated pressure or vacuum losses across the piping as designed do not exceed ____ percent of the intended operating pressure or vacuum at the source valve.

a)

10

b)

15

c)

20

d)

25

78.

The filter shall accrue no more than .001g (__ mg) of matter from any outlet tested.

a)

.001

b)

.01

c)

.10

d)

1.0

79.

A standard gauge pressure of 0-300 psi, applies to which of the following?

a)

Instrument air

b)

Laboratory air

c)

Non-medical air

d)

Laboratory vacuum

80.

Stainless steel tube may be joined by any of the following methods EXCEPT which of the following?

a)

Tungsten Inert Gas (TIG)

b)

Metal Inert Gas (MIG)

c)

Solder

d)

Gas Tungsten Arc Welding (GTAW)

81.

The gas produced by evaporation from the secondary header in a cryogenic liquid container system, shall enter the supply system _________ of the final line regulator.

a)

downstream

b)

upstream

c)

upstream of the filter

d)

not required

82.

Brazed joints shall be made using a brazing alloy that exhibits a melting temperature in excess of ________° F, to retain the integrity of the piping system in the event of fire.

a)

800

b)

1000

c)

1200

d)

2400

83.

In a facility where medical gas or vacuum systems are present and the failure is likely to cause imminent major injury or death to patients or caregivers, the system shall be designed to meet which category of requirements?

a)

1

b)

2

c)

3

d)

4

84.

Cylinders in use and in storage shall be prevented from reaching temperatures in excess of _______° F.

a)

120

b)

125

c)

130

d)

150

85.

Which of these is a requirement for each station inlet?

a)

A main vacuum line

b)

A primary optional

in-line

c)

A primary valve

d)

A secondary demand valve

86.

Which of the following is true of Category 1 support gases?

a)

Permitted to power medical-surgical tools

b)

Subject to greater hazards than present in other piped gases

c)

Used for respiration

d)

Permitted to power medical-surgical tools and subject to greater hazards than present in other piped gases

87.

WAGD networks shall provide a WAGD inlet in all locations where _______ is intended to be administered.

a)

Medical air

b)

Nitrous oxide

c)

Oxygen

d)

Nitrogen

88.

A bulk nitrous oxide system has a storage capacity of more than ________ pounds of nitrous oxide.

a)

20,000

b)

3,200

c)

16,000

d)

5,000

89.

NFPA 99 requires tubing to be cut square using a sharp tubing cutter, that shall be free from grease, oil or other lubricants not suitable for oxygen service. What sort of treatment is permitted of the tube ends before brazing?

a)

May be reamed

b)

Shall be deburred or rolled smooth

c)

may be re-cleaned

d)

none that are listed are permitted

90.

Category 1 facility testing and calibration procedures shall be established, to ensure carbon monoxide monitors are calibrated _______________.

a)

annually or more often, if recommended by the manufacturer

b)

as directed by the administrator of the facility

c)

at least once, every 18 months

d)

by an independent calibration/testing company

91.

All tools for medical gas work shall be kept separate from other tools, and kept free of grease or oil. If cleaned, only a degreaser such as ___________ is permitted.

a)

a natural citrus solvent cleaner-degreaser

b)

water

c)

trisodium phosphate or sodium carbonate

d)

Goop or GoJo hand cleaner

92.

Manufactured assemblies employing flexible hose shall NOT be labeled with ____________.

a)

system name or symbol

b)

color code

c)

standard pressure

d)

date of installation

93.

Bulk cryogenic liquid systems shall be installed in accordance with ________

a)

NFPA 55

b)

NFPA 45

c)

CGA 55

d)

CGA 50

94.

When is the initial pressure test conducted by the installer?

a)

After installation of alarm sensors

b)

Before blow down of distribution piping

c)

Before installation of station rough-in assemblies

d)

After blow down of distribution piping

95.

What is the test pressure for the initial pressure test conducted by the installer?

a)

1.5 times the normal operating pressure

b)

50 cfm

c)

500 cfm

d)

1.5 times the normal operating pressure, but not less than a gauge pressure of 150 psi

96.

Joints shall be brazed within ______ hours after the surfaces are cleaned for brazing.

a)

one

b)

four

c)

eight

d)

twenty-four

97.

NFPA 99 code, states that medical air may be supplied from all EXCEPT which of the following?

a)

Instrument air compressors

b)

Bulk containers

c)

Medical air compressors

d)

Reconstituted from oxygen USP and oil-free, dry nitrogen NF

98.

Which type of connectors shall connect the medical air compressors with their intake and outlet piping?

a)

Insulated

b)

Compression

c)

Flexible

d)

Dielectric

99.

Underground piping buried outside of buildings, shall be installed below the local level of frost penetration, and shall be protected by a(n) __________________.

a)

conduit

b)

enclosure

c)

cover

d)

any of the listed protections are acceptable

100.

Performance qualification of brazers shall remain in effect indefinitely, unless the brazer does NOT braze with the qualified procedure for longer than what period of time?

a)

6 months

b)

12 months

c)

2 years

d)

Brazing certs do not expire