NEW
Font size
WorksheetsImag II: Final "Mini Quiz Me"
Total questions: 77
Worksheet time: 39mins
The purpose of filtration is to:
Increase patient dose
Remove low-energy photons
Increase image contrast
Decrease HVL
The Half-Value Layer (HVL) measures:
Beam quantity
Beam intensity at the anode
Beam quality/penetration
Scatter level in the room
Adding filtration will:
Decrease patient skin dose
Increase the number of soft photons
Increase beam intensity
Decrease beam hardness
Increasing HVL means the beam:
Has lower penetration
Contains more low-energy photons
Is more penetrating
Has reduced kVp
Inherent filtration is found:
Inside the tube housing
Between the patient and IR
Outside the collimator
In the AEC detector
Which is the primary material used for filtration?
Lead
Copper
Aluminum
Silver
Added filtration is most commonly contributed by:
Anode
Window of tube
Collimator mirror
AEC ion chambers
Beam hardening refers to:
Increasing mAs
Increasing the average photon energy
Reducing tube cooling time
Increasing scatter production
As filtration increases, IR exposure:
Increases
Decreases
Stays the same
Doubles
Low-energy photons should be removed because they:
Improve contrast only
Cause unnecessary patient dose
Do not interact with tissue
Always scatter away from patient
AEC works by terminating exposure based on:
Exposure time
Amount of radiation reaching the detector
mAs selected
Detector temperature
Mispositioning over the wrong AEC cell will result in:
Perfect exposure
Consistent density
Overexposure or underexposure
Increased contrast
If the anatomy does not fully cover the AEC cell, the image will likely be:
Too dark
Too light
Perfect
Too magnified
Collimating too tightly when using AEC may cause:
Scatter increase
Exposure termination too early
Longer exposure times
Saturation of the IR
The backup time exists to:
Increase patient dose
Prevent overexposure
Increase tube heating
Reduce contrast
Increasing mA when using AEC will:
Increase exposure time
Decrease exposure time
Not affect exposure time
Make the image darker
The density (exposure) control on AEC adjusts:
kVp
Beam quality
Exposure termination point
Detector size
Grids are used primarily to:
Reduce patient dose
Remove scatter before reaching the IR
Increase beam intensity
Reduce field size
Scatter radiation primarily affects image:
Contrast
Sharpness
Brightness
Resolution
Compton scatter is more likely to occur in:
Bone
Metal
Soft tissue
Air
Decreasing field size will:
Increase scatter
Decrease scatter
Increase patient dose
Increase fog
The most effective way to reduce scatter is:
Increase SID
Reduce OID
Collimate tightly
Increase kVp
Beam restriction improves image:
Density
Contrast
Distortion
Magnification
Off-level grid cutoff appears as:
One side light
Both sides light
Overall underexposure
Only the center exposed
Off-center grid cutoff appears as:
One-sided cutoff
Complete loss of exposure
Off-focus grid cutoff happens when:
Wrong grid ratio is selected
SID is outside focal range
Tube is angled crosswise
Detector is rotated
Upside-down focused grid results in:
Centered exposure only
Severe cutoff on both edges
Dark edges
Increased density
A grid should be used when the body part is thicker than:
3 cm
5 cm
10 cm
20 cm
A higher grid ratio will:
Decrease contrast
Require more mAs
Increase scatter reaching IR
Reduce patient dose
Air gap technique reduces scatter by:
Increasing distance between patient and IR
Increasing filtration
Reducing OID
Increasing kVp
The interaction responsible for scatter is:
Photoelectric absorption
Bremsstrahlung
Characteristic
Photoelectric absorption increases:
Fog
Patient dose
Scatter
Focal spot size
Increasing kVp will:
Increase scatter
Decrease scatter
Increase grid ratio
Reduce beam penetration
The 15% rule states that increasing kVp by 15% will:
Double IR exposure
Halve IR exposure
Reduce contrast
Increase patient dose only
mAs controls:
Beam energy
Beam quantity
HVL
Scatter production
Increasing mAs will:
Increase contrast
Increase IR exposure
Decrease patient dose
Increase penetration
When SID increases, IR exposure:
Increases
Decreases
Doubles
Bremsstrahlung photons are created when electrons:
Knock out inner-shell electrons
Slow down near the nucleus
Combine with protons
Produce scatter
Characteristic photons occur when:
Electrons change direction
Outer-shell electrons fill inner-shell vacancies
Electrons pass near the nucleus
Electrons strike the filament
Most x-ray photons produced at 70 kVp are:
Photoelectric
Characteristic
Compton
Bremsstrahlung
Increasing SID will:
Increase magnification
Decrease magnification
Increase scatter
Increase contrast
Decreasing OID will:
Increase magnification
Decrease magnification
Increase scatter
Reduce focal spot blur
Increasing kVp affects photoelectric absorption by:
Increasing PE
Decreasing PE
Not affecting PE
Best way to reduce patient dose:
Increase mAs
Increase SID
Use highest kVp appropriate
Use large field size
Increasing collimation (smaller field size) will:
Increase scatter
Reduce scatter
Increase magnification
Increase patient dose
AEC minimum response time may cause:
Overexposure
Underexposure
Motion blur
Increased contrast
When using AEC, which factor must still be selected manually?
Exposure time
mA
Termination point
Density setting
If the patient is thinner than expected, AEC will:
Overexpose the IR
Underexpose the IR
Produce perfect density
Increase contrast
If SID increases from 40" to 80", exposure will become:
4x greater
½ as intense
¼ as intense
Unchanged
Original technique is 10 mAs at 40". What mAs is needed at 56"? Use exposure maintenance formula.
8 mAs
10 mAs
16 mAs
20 mAs
If kVp is increased by 15%, mAs must be:
Doubled to maintain density
Halved to maintain density
Reduced by 25%
Increased by 30%
If mA = 200 and time = 0.1 s, what is mAs?
2
10
20
200
Doubling mAs will:
Double IR exposure
Half IR exposure
Quadruple exposure
Reduce scatter
Increasing filtration will:
Increase contrast
Decrease patient skin dose
Increase density
Increase Compton scatter
Total filtration includes:
Only inherent
Only added
Inherent + added
Compensating filters only
Compensating filters are used to:
Reduce scatter
Even out tissue densities
Increase contrast
Increase beam penetration
Wedge filters are useful for:
Knee
Skull
AP foot
Chest
The primary benefit of beam restriction is:
Increase kVp
Reduce scatter
Increase spatial resolution
Increase mAs
The collimator mirror adds:
Inherent filtration
Added filtration
Compensating filtration
Scatter filtration
A light field that does NOT match the x-ray field violates:
AEC
SID
Beam alignment
HVL
Off-level grid occurs when:
SID is too short
Tube is tilted
Wrong grid ratio
Patient is rotated
Using a higher grid ratio requires:
an increase in exposure
a decrease in exposure
no change in exposure
lower patient dose
Bremsstrahlung production depends on:
Binding energy
Distance to nucleus
mAs only
Electron capture
Characteristic photons depend on:
kVp above 69 keV
Filament temperature
Tube cooling
OID
The majority of beam interactions in tissue are:
Pair production
Compton scatter
Photoelectric
Coherent scatter
Coherent scatter has:
Major impact on image
No energy transfer
High patient dose
High penetration
Increasing collimation will:
Increase fog
Decrease fog
Increase contrast loss
Increase Compton scatter
If the patient is too wide for the AEC cell, the result may be:
More mAs
Less mAs
Higher SID
Lower kVp
What is the result of using too high mAs?
Overexposure
Underexposure
Perfect exposure
Increased contrast
Backup timer is typically set at:
50% of expected mAs
100% of expected mAs
150% of expected mAs
200% of expected mAs
A high kVp technique results in:
More PE
More Compton
Less scatter
Lower HVL
The anode heel effect is more noticeable with:
Short SID
Large focal spot
Small field size
High filtration
Increasing filtration affects the beam by:
Decreasing quality
Increasing quality
Removing high-energy photons
Increasing Compton scatter
The purpose of collimator shutters is to:
Reduce beam quality
Reduce off-focus radiation
Increase beam intensity
Increase mAs
74. HVL increases when:
the energy of the X-ray beam increases
the energy of the X-ray beam decreases
the filtration is reduced
the tube current is increased
Which of the following occurs when the grid ratio increases?
Tube warms
Filtration increases
mAs increases
Grid ratio increases
The most effective scatter cleanup device is:
Aluminum filter
Air gap
Collimator
High-ratio grid
